Oxygen Tubing - Salter Labs
of the tubing should be no less than 3/16" (0.48 cm):
the outlet port of the humidifier or directly to the oxygen outlet port if you do not use a
humidifier. Connect the other end of the tube to the nasal cannula, if not already attached.
Tubing not specified for use with this oxygen system may impair the performance.
The tube fitting may be tight. DO NOT use oil or grease to lubricate.
WARNING
If tubing becomes disconnected during operation, ECLIPSE does not
WARNING
alarm. Check flow at the cannula to verify oxygen delivery.
To ensure there is adequate flow to deliver oxygen, the length of the
CAUTION
tubing must not exceed 50 feet (15.2m) when using Continuous Flow
Mode, and must not exceed 7 feet (2.1m) when operating in Pulse
Flow Mode for adequate breath detection.
Cannula – Salter Labs
have prescribed a cannula to deliver oxygen. In most cases they are already attached
to the oxygen tubing. If not, follow the instructions included with the cannula to attach
it to the oxygen tubing. Use of cannula not specified for use with this oxygen system
may impair the performance.
You have various options to provide power to your ECLIPSE.
Power Options:
A Power Cartridge, AC Power Supply and DC Power Supply.
Power Cartridge, Model 2400:
your ECLIPSE is capable of providing you with battery operation depending on your
prescribed flow rate.
Your Power Cartridge will be recharged when the ECLIPSE is connected to the AC
Power Supply or the DC Power Supply, if adequate power is available. The Power
Cartridge will take longer to recharge when connected to the DC Power Supply and
will depend on the flow setting and the amperage switch setting.
Power Compartment:
operation is not required. When using the Power Compartment, your ECLIPSE will
ONLY operate on the AC and DC Power Supplies.
®
Series 2000 or equivalent tubing (internal diameter
®
Part Number 1600 Series or equivalent:
The rechargeable Power Cartridge supplied with
The Power Compartment is available when Power Cartridge
Model 1000 User Manual
Connect the oxygen tubing to
Your physician will
ECLIPSE
TM
Model 1000
User Manual
| 33