Symptom
ECLIPSE does not power on
when ON/OFF button is pressed
No Oxygen
Low Oxygen Concentration
Low Oxygen Flow
No Oxygen delivered in
Pulse Mode
Power Cartridge Status Gauge
never indicates fully charged
Probable Cause
No Power Cartridge installed
Power Cartridge is discharged
Malfunction
Blocked or kinked tubing
Filter Blocked
Humidifier
ECLIPSE not ON
Tubing or cannula is not properly
connected or kinked
Power Cartridge is discharged
or no external power is present
Malfunction
Restriction in tubing or humidifier Repair or replace tubing
Filter restricted
Malfunction
Restriction in humidifier or tubing Repair or replace humidifier
Filter Blocked
Malfunction
Tubing/cannula longer than
7 feet (2.1m)
Humidifier attached
No inspiration detected
Power Cartridge needs
calibration
Model 1000 User Manual
Remedy
Install charged Power Cartridge
or plug into external power
Plug into external power
Contact your dealer/provider
Check tubing or replace
Clean Air Inlet Filter
Check humidifier attachment
and tubing
Turn ECLIPSE ON
Check tubing, cannula and
connections
Install charged Power Cartridge
or attach external power
Contact your dealer/provider
or humidifier
Clean or replace air inlet filter
Contact your dealer/provider
or tubing
Clean or replace air inlet filter
Contact your dealer/provider
Attach 7 foot (2.1m)
tubing/cannula
Remove humidifier
Contact your dealer/provider
Fully discharge the Power
Cartridge and recharge
ECLIPSE