Typical Power Cartridge Recharge Time
The typical time to recharge your Power Cartridge to achieve 80% capacity from
a fully discharged Power Cartridge is 1.4 hours to 5.0 hours, dependent upon the
flow setting.
If the Power Cartridge is too warm, charging will not begin until it sufficiently cools.
The Power Cartridge may not charge or the charge time may increase when operating
from the DC Power Supply set in the 10 AMP position.
I N I T I A L P OW E R CA R T R I D G E O P E R AT I O N :
The new Power Cartridge packaged with your ECLIPSE Oxygen System is not fully
charged. Before using your ECLIPSE Oxygen System for the first time, you must fully
discharge the Power Cartridge.
Turn your ECLIPSE ON and allow the Power Cartridge to completely discharge.
1.
Unit will sound a low Power Cartridge alert until the unit completely shuts down.
This could continue for up to approximately 20 minutes.
Connect the AC Power Supply and plug in the AC Power Cord to completely
2.
charge the Power Cartridge. This may require up to five hours of uninterrupted
charging. This completes the calibration of the Power Cartridge.