Technical Data

3. Technical Data

3.1 Specifications

3.1.1 TIP900SYS (Thermal Imager, needs to be mounted in protective housing)

Type
Temperature Range
Spectral Response
Detector
System Accuracy
Repeatability
Temperature Resolution
Scan Rate
Measured Points per Line (at FOV)
Scan Angle (FOV)
Standard Focal Distance
Optical Resolution D:S (90% energy) 170 : 1 (IFOV = 5.9 mrad)
Hot Spot Detection (50% energy)
Mechanical scanning system
Power Supply
Power Supply to protective housing
Protection rate for MP150LT
Protection rate for protective housing IP44 (NEMA3), Rittal cabinet
Ambient Operation Temperatures at still air (e.g. MP150LT integrated in protective housing):
- Without cooling
Internal Operation Temperatures
- Laser
Storage Temperatures
Air Purge Collar: Pressure
Air Purge Collar: Flow rate
Warm-up Time
Relative Humidity
Vibration
Shock
Dimensions
Weight (incl. air purge)
16
Raytek MP150LT
20 to 350°C (68 to 662°F)
3 to 5 µm
MCT
± 2°C (± 4°F) at 0 - 50°C (32 - 122°F) ambient temperature
± 1°C (± 2°F) at 0 - 50°C (32 - 122°F) ambient temperature
0.1 K (digital interface), 16 bit (analog output)
20 Hz, 36 Hz, 48 Hz, 76 Hz, 85 Hz, 108 Hz, 126 Hz, 150 Hz
256 pixel @ 150 Hz, 512 pixel @ 76 Hz, 1024 pixel @ 36 Hz
90°
1.52 m (60")
510 : 1 (IFOV = 2 mrad)
MTBF: 40,000 hours at 48Hz scan rate
24 VDC ± 25%, 1A
100 to 240 VAC, 50/60Hz
IP65 (NEMA4) in accordance to IEC 60529
0 to 50°C (32 to 122°F), no direct sunlight
0 to 60°C (32 to 140°F)
automatic switch off at < 5°C (41°F) or > 50°C (122 °F)
-25 to 65°C (-13 to 149°F)
max. 3 bar (43 psig)
typ. 100 l/min to 200 l/min (3.53 to 7.06 cfm)
30 min.
10 to 90%, non-condensing, for operating and non-operating
IEC 60068-2-6, 3 axes, 10 to 150 Hz, operating 2 g above 20
Hz
IEC 60068-2-27, 3 axes, operating: 5 g at 11 ms, 15 g at 6 ms
200 x 180 x 190 mm (7.8 x 7.02 x 7.41 in)
7 kg (15.4 lbs)