4.3.2 Electrical Installation of the Protective Housing

Figure 25: Principle diagram of internal wiring in Protective Housing

The external wiring to the Protective Housing consists in total of four cables, which are:
100 – 240VAC power supply, ~ 100W, max. 2.5mm
24VDC, ~ 6W power supply to the Temperature Compensation Network Box, max. 0.5mm
AWG21, shielded cable
RJ45 LAN to the Temperature Compensation Network Box (< 80m distance), min. Cat5e
RJ45 LAN to the Industrial Touch-Panel PC (< 80m distance), min Cat5e
If installing the TIP900 Protective Housing in a warm environment, water-cooling may be
necessary. The tubes used for water may be run through the second grommet plate!