Field Communicator application
September 2021
3.15
3.15.1
3.15.2
94
1.800.561.8187
Offline HART configurations
The Trex unit lets you work with HART configurations. You can create, edit, copy, delete,
and rename configurations.
If you are not connected to a HART device, an Offline Config button displays on the main
Field Communicator menu. You can edit and store a configuration on the Trex unit.
If you are connected to a HART device, you can either Save Config to the Trex unit, or Send
Config to the connected device.
You can also back up the configurations to a PC by using the Configuration Management
application. For more information, see the Help included with the Configuration
Management application.
Note
Do not edit a configuration on the Trex unit while you are connected to the Configuration
Management PC application.
Create a HART configuration
Create a new configuration for a specific device type and revision. You can then apply this
configuration to multiple devices.
Procedure
1. In the Field Communicator application, tap HART Offline.
2. Tap Offline Config → New.
3. Tap a manufacturer.
4. Tap a device type.
5. Tap a device revision.
6. Tap OK.
7. Select and edit the necessary parameters.
8. Tap OK to mark the parameter for writing.
Note
Unmarked parameters are saved but not sent when you transfer the configuration
to a connected device.
9. Tap Save when you finish editing the parameters.
10. Enter a name for the configuration, and tap OK.
The configuration file is saved.
Edit a HART configuration
Prerequisites
Note
Do not edit a configuration on the Trex unit while you are connected to the Configuration
Management PC application.