Emerson AMS Trex Device Communicator User Manual page 17

Table of Contents

Advertisement

User Guide
2.5.2
User Guide
1.800.561.8187
WARNING
Do not use the HART + pwr and mA terminals in series when the Trex unit powers a
device.
Do not connect the mA terminals (ammeter) in parallel with a powered 4-20 mA
current loop. Ammeters have low resistance. This can disrupt the loop and cause
devices to report incorrect values or positioners to move unexpectedly.
Do not connect the mA terminals on the Trex unit to a power supply that is not current
limited to 250 mA. This can blow a fuse inside the Trex unit. The repair/replacement
will need to be completed at an authorized service center.
For best results when using the mA terminals, do the following:
Avoid using the ammeter in cold or very hot temperatures. Extreme temperatures can
affect the accuracy of the measurements.
Connect the mA terminals on the Trex unit to the "low-side" of the device. See the
example below.
Figure 2-6: "Low-side" ammeter connection
A. Analog output
Install a communication module
CAUTION
Ensure sufficient grounding. Ensure the personnel, working surfaces, and packaging
are sufficiently grounded when handling electrostatically sensitive parts.
Avoid touching the pins on the connectors or components. Discharged energy can
affect the modules.
Handle the module with care. It is possible to bend the pins on the module's connector.
www.
AMS Trex Device Communicator overview
information@itm.com
.com
September 2021
17