Field Communicator application
September 2021
78
1.800.561.8187
Figure 3-20: Communicate with the THUM adapter attached to a powered HART
transmitter
Connect the lead set to the yellow and white wires.
Note
If you want to communicate with the THUM adapter, ensure the Auto-Connect option for
the Field Communicator Settings is disabled. This allows the Field Communicator
application to detect both the THUM adapter and the transmitter, and let you select which
one to communicate with.
A. Optional connection if the yellow and white wires are inaccessible.
B. Voltage source.
Device connection wizard prompts
Do you intend to provide power from the Trex unit?
Leads are attached to:
Change Polling Option?
www.
Answer
No
Transmitter
Tap Yes and select the address for the
THUM adapter, the device, or "0,63"
for both.