Figure 3-19: Power and communicate with the THUM adapter attached to a HART
transmitter
Note
If you want to communicate with the THUM adapter rather than the device, ensure the
Auto-Connect option for the Field Communicator Settings is disabled. This allows the Field
Communicator application to detect both the THUM adapter and the transmitter, and let
you select which one to communicate with.
Device connection wizard prompts
Do you intend to provide power from the Trex unit?
Are you providing power to a transmitter or positioner?
Change Polling Option?
www.
Field Communicator application
Answer
Yes
Transmitter
Tap Yes and select the address for the
THUM adapter, the device, or "0,63"
for both.