8.3.1.2

Install a new display

Note
Mounting orientation for display
The display can be mounted in 1 of 4 positions at 90 degree intervals, for easy viewing after
installation.
1. Press the female end of cable from replacement display onto male four-pin connector.
2. Set replacement display into enclosure approximately one quarter turn counter-clockwise
from the desired final orientation. Gently turn display one quarter turn clockwise to secure
the display in the enclosure.
3. Replace device lid. Thread onto enclosure, turning clockwise. Hand tighten until mechanical
stop is reached.
8.4

Return procedure

Enclose the bill of lading, return document and decontamination certificate in a clear plastic
pouch and attach it firmly to the outside of the packaging. Any devices/replacement parts
which are returned without a decontamination declaration will be cleaned at your expense
before further processing. For further details, refer to the operating instructions.
SITRANS Probe LU240 with mA/HART
Compact Operating Instructions, 09/2021, A5E42673704-AE
Male four-pin connector
Female end of display cable
Service and maintenance
8.4 Return procedure