Replacing Meter Body; Figure 27 - Disassembly For Meter Body Replacement; Figure 28 - Hardware Location To Remove The Meter Assembly - Honeywell SMV800 SmartLine User Manual

Multivariable transmitter
Hide thumbs Also See for SMV800 SmartLine:
Table of Contents

Advertisement

10.5 Replacing Meter Body

Meter body can be replaced by following the steps below.
1. Save or record device configuration data.
2. Turn off Transmitter power.
3. Remove Transmitter from service and place it in a clean area before disassembling.
4. Refer to Figure 27. Loosen End Cap Lock and unscrew End Cap from the electronics side of
Electronics Housing.
Take appropriate steps to avoid ESD damage when handling Communication and
Display Modules.
5. If Display Module is present, press the two snaps on the side to remove it from Communication
Module.
Note: Do not discard or misplace the connector which connects Display Module and
Communication Module, it will be required tin reassembling
6. Loosen the two retaining screws to remove Communication Module from Electronics Housing.
7. Disconnect Sensor Cable Connector from Communications Module.
8. Refer to Figure
allow rotating the complete meter body.
Page 68
Figure 27 – Disassembly for Meter Body Replacement
28.
Use a 2 mm hex wrench to loosen Set Screw on the outside of the housing to
Figure 28 – Hardware Location to Remove the Meter Assembly