Endress+Hauser Proline Promass O 500 PROFIBUS DP Operating Instructions Manual

Endress+Hauser Proline Promass O 500 PROFIBUS DP Operating Instructions Manual

Coriolis flowmeter
Table of Contents

Advertisement

Quick Links

BA01876D/06/EN/01.18
71402571
Valid as of version
01.00.zz (Device firmware)
Products
Operating Instructions
Proline Promass O 500
PROFIBUS DP
Coriolis flowmeter
Solutions
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Promass O 500 PROFIBUS DP

  • Page 1 Products Solutions Services BA01876D/06/EN/01.18 71402571 Valid as of version 01.00.zz (Device firmware) Operating Instructions Proline Promass O 500 PROFIBUS DP Coriolis flowmeter...
  • Page 2 • The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these instructions. Endress+Hauser...
  • Page 3: Table Of Contents

    Proline Promass O 500 PROFIBUS DP Table of contents Table of contents About this document ....6 Installation ..... . . 23 Document function .
  • Page 4 Table of contents Proline Promass O 500 PROFIBUS DP Access to the operating menu via the local Commissioning ....105 display ......
  • Page 5 13.2 Measuring and test equipment ... 244 13.3 Endress+Hauser services ....244 Repairs ......245 14.1 General notes .
  • Page 6: About This Document

    About this document Proline Promass O 500 PROFIBUS DP About this document Document function These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
  • Page 7: Tool Symbols

    Proline Promass O 500 PROFIBUS DP About this document Symbol Meaning Light emitting diode is on. Light emitting diode is flashing. 1.2.4 Tool symbols Symbol Meaning Torx screwdriver Phillips head screwdriver Open-ended wrench 1.2.5 Symbols for certain types of information...
  • Page 8: Documentation

    • W@M Device Viewer (www.endress.com/deviceviewer): Enter the serial number from nameplate • Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2D matrix code (QR code) on the nameplate Detailed list of the individual documents along with the documentation code →...
  • Page 9 Proline Promass O 500 PROFIBUS DP About this document TRI-CLAMP® Registered trademark of Ladish & Co., Inc., Kenosha, USA Endress+Hauser...
  • Page 10: Basic Safety Instructions

    Basic safety instructions Proline Promass O 500 PROFIBUS DP Basic safety instructions Requirements for the personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
  • Page 11: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 12: Product Safety

    It meets general safety standards and legal requirements. It also complies with the EU directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser confirms this by affixing the CE mark to the device.
  • Page 13: Protecting Access Via A Password

    Proline Promass O 500 PROFIBUS DP Basic safety instructions 2.7.2 Protecting access via a password Different passwords are available to protect write access to the device parameters or access to the device via the WLAN interface. • User-specific access code Protect write access to the device parameters via the local display, Web browser or operating tool (e.g.
  • Page 14: Access Via Service Interface (Cdi-Rj45)

    Basic safety instructions Proline Promass O 500 PROFIBUS DP The device and status information can be hidden on the login page. This prevents unauthorized access to the information. For detailed information on device parameters, see: The "Description of Device Parameters" document →  278.
  • Page 15: Product Description

    Proline Promass O 500 PROFIBUS DP Product description Product description The measuring system consists of a transmitter and a sensor. The transmitter and sensor are mounted in physically separate locations. They are interconnected by connecting cables. Product design Two versions of the transmitter are available.
  • Page 16 Product description Proline Promass O 500 PROFIBUS DP 3.1.2 Proline 500 Signal transmission: analog Order code for "Integrated ISEM electronics", option B "Transmitter" For use in applications required to meet special requirements due to ambient or operating conditions. As the electronics are located in the transmitter, the device is ideal in the event of: •...
  • Page 17: Incoming Acceptance And Product

    Technical Documentation present? • If one of the conditions is not satisfied, contact your Endress+Hauser Sales Center. • Depending on the device version, the CD-ROM might not be part of the delivery! The Technical Documentation is available via the Internet or via the Endress+Hauser Operations App, see the "Product identification"...
  • Page 18: Transmitter Nameplate

     8 sections • The W@M Device Viewer: enter the serial number from the nameplate (www.endress.com/deviceviewer) • The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate. 4.2.1 Transmitter nameplate Proline 500 –...
  • Page 19 Proline Promass O 500 PROFIBUS DP Incoming acceptance and product identification Proline 500 3 4 5 Order code: Ser. no.: Ext. ord. cd.: Date: A0029192  4 Example of a transmitter nameplate Manufacturing location Name of the transmitter Order code Serial number (ser.
  • Page 20: Sensor Nameplate

    Incoming acceptance and product identification Proline Promass O 500 PROFIBUS DP 4.2.2 Sensor nameplate Order code: Ser. no.: Ext. ord. cd.: Date: A0029199  5 Example of a sensor nameplate Name of the sensor Manufacturing location Order code Serial number (ser. no.) Extended order code (Ext.
  • Page 21: Symbols On Measuring Device

    Proline Promass O 500 PROFIBUS DP Incoming acceptance and product identification 4.2.3 Symbols on measuring device Symbol Meaning WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. Reference to documentation Refers to the corresponding device documentation.
  • Page 22: Storage And Transport

    Storage and transport Proline Promass O 500 PROFIBUS DP Storage and transport Storage conditions Observe the following notes for storage: ‣ Store in the original packaging to ensure protection from shock. ‣ Do not remove protective covers or protective caps installed on process connections.
  • Page 23: Measuring Devices With Lifting Lugs

    Proline Promass O 500 PROFIBUS DP Installation 5.2.2 Measuring devices with lifting lugs CAUTION Special transportation instructions for devices with lifting lugs ‣ Only use the lifting lugs fitted on the device or flanges to transport the device. ‣ The device must always be secured at two lifting lugs at least.
  • Page 24 Installation Proline Promass O 500 PROFIBUS DP To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the following mounting locations in the pipe: • Highest point of a pipeline. • Directly upstream of a free pipe outlet in a down pipe.
  • Page 25 Proline Promass O 500 PROFIBUS DP Installation Orientation Recommendation Horizontal orientation, transmitter at bottom Exceptions: →  7,  25 A0015590 Horizontal orientation, transmitter at side A0015592 Applications with low process temperatures may decrease the ambient temperature. To maintain the minimum ambient temperature for the transmitter, this orientation is recommended.
  • Page 26: Requirements From Environment And Process

    If operating outdoors: Avoid direct sunlight, particularly in warm climatic regions. You can order a weather protection cover from Endress+Hauser. →  247. System pressure It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas.
  • Page 27 Proline Promass O 500 PROFIBUS DP Installation NOTICE Electronics overheating on account of thermal insulation! ‣ Recommended orientation: horizontal orientation, sensor connection housing pointing downwards. ‣ Do not insulate the sensor connection housing. ‣ Maximum permissible temperature at the lower end of the sensor connection housing: 80 °C (176 °F)
  • Page 28: Special Mounting Instructions

    Installation Proline Promass O 500 PROFIBUS DP Vibrations The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by plant vibrations. 6.1.3 Special mounting instructions Rupture disk Information that is relevant to the process: →  266.
  • Page 29 Proline Promass O 500 PROFIBUS DP Installation Experience shows that zero point adjustment is advisable only in special cases: • To achieve maximum measuring accuracy even with low flow rates. • Under extreme process or operating conditions (e.g. very high process temperatures or very high-viscosity fluids).
  • Page 30: Mounting The Measuring Device

    Installation Proline Promass O 500 PROFIBUS DP SW 2.5 5 (0.2) min. 15 (0.6) A0029799 Cover borehole for the securing screw Securing screw to lock the cover Mounting the measuring device 6.2.1 Required tools For transmitter For mounting on a post: •...
  • Page 31: Mounting The Transmitter Housing: Proline 500 - Digital

    Proline Promass O 500 PROFIBUS DP Installation 2. Install the measuring device or turn the transmitter housing so that the cable entries do not point upwards.  A0029263 6.2.4 Mounting the transmitter housing: Proline 500 – digital CAUTION Ambient temperature too high! Danger of electronics overheating and housing deformation.
  • Page 32: Mounting The Transmitter Housing: Proline 500

    Installation Proline Promass O 500 PROFIBUS DP Wall mounting 17 (0.67) 5.8 (0.23) 5.8 (0.23) 149 (5.85) A0029054  12 Engineering unit mm (in) Depends on order code for "Transmitter housing" Order code for "Transmitter housing" • Option A, aluminum coated: L =14 mm (0.55 in) •...
  • Page 33 Proline Promass O 500 PROFIBUS DP Installation Wall mounting 18 (0.71) 10 (0.39) 20 (0.79) 100 (3.94) A0029068  13 Engineering unit mm (in) 1. Drill the holes. 2. Insert wall plugs into the drilled holes. 3. Screw in the securing screws slightly at first.
  • Page 34: Turning The Transmitter Housing: Proline 500

    Installation Proline Promass O 500 PROFIBUS DP Post mounting WARNING Order code for "Transmitter housing", option L "Cast, stainless": cast transmitters are very heavy. They are unstable if they are not mounted on a secure, fixed post. ‣ Only mount the transmitter on a secure, fixed post on a stable surface.
  • Page 35: Turning The Display Module: Proline 500

    Proline Promass O 500 PROFIBUS DP Installation 7. Depending on the device version: Attach the securing clamp of the connection compartment cover. 6.2.7 Turning the display module: Proline 500 The display module can be turned to optimize display readability and operability.
  • Page 36: Electrical Connection

    Electrical connection Proline Promass O 500 PROFIBUS DP Electrical connection NOTICE The measuring device does not have an internal circuit breaker. ‣ For this reason, assign the measuring device a switch or power-circuit breaker so that the power supply line can be easily disconnected from the mains.
  • Page 37 Proline Promass O 500 PROFIBUS DP Electrical connection Signal damping Max. 9 dB over the entire length of the cable cross-section Shield Copper braided shielding or braided shielding with foil shield. When grounding the cable shield, observe the grounding concept of the plant.
  • Page 38 Electrical connection Proline Promass O 500 PROFIBUS DP A0032476 Proline 500 digital transmitter Proline 500 transmitter Sensor Promass Non-hazardous area Hazardous area: Zone 2; Class I, Division 2 Hazardous area: Zone 1; Class I, Division 1 Standard cable to 500 digital transmitter →  38 Transmitter installed in the non-hazardous area or hazardous area: Zone 2;...
  • Page 39 Proline Promass O 500 PROFIBUS DP Electrical connection Optionally available connecting cable Design 2 × 2 × 0.34 mm (AWG 22) PVC cable with common shield (2 pairs, uninsulated stranded CU wires; pair-stranded) Flame resistance According to DIN EN 60332-1-2...
  • Page 40 Electrical connection Proline Promass O 500 PROFIBUS DP Cross-section Cable length [max.] Termination 2 x 2 x 0.50 mm 50 m (165 ft) 2 x 2 x 0.50 mm (AWG 20) (AWG 20) BN WT YE GN • +, – = 0.5 mm •...
  • Page 41: Terminal Assignment

    Proline Promass O 500 PROFIBUS DP Electrical connection C: Connecting cable between sensor and transmitter: Proline 500 Standard cable 6 × 0.38 mm PVC cable with common shield and individually shielded cores Conductor resistance ≤50 Ω/km (0.015 Ω/ft) Capacitance: core/shield ≤420 pF/m (128 pF/ft)
  • Page 42: Preparing The Measuring Device

    Electrical connection Proline Promass O 500 PROFIBUS DP Where applicable, national installation regulations and guidelines must be observed during the installation! Where there are large differences in potential between the individual grounding points, only one point of the shielding is connected directly with the reference ground. In systems without potential equalization, therefore, cable shielding of fieldbus systems should only be grounded on one side, for example at the fieldbus supply unit or at safety barriers.
  • Page 43 Proline Promass O 500 PROFIBUS DP Electrical connection 2. If the measuring device is supplied without cable glands: Provide suitable cable gland for corresponding connecting cable. 3. If the measuring device is supplied with cable glands: Observe requirements for connecting cables →  36.
  • Page 44: Connecting The Measuring Device: Proline 500 - Digital

    Electrical connection Proline Promass O 500 PROFIBUS DP Connecting the measuring device: Proline 500 - digital NOTICE Limitation of electrical safety due to incorrect connection! ‣ Have electrical connection work carried out by appropriately trained specialists only. ‣ Observe applicable federal/national installation codes and regulations.
  • Page 45 Proline Promass O 500 PROFIBUS DP Electrical connection Connecting the sensor connection housing via terminals For the device version with the order code for "Sensor connection housing": • Option A "Aluminum coated" • Option L "Cast, stainless" 10 (0.4) 3 mm...
  • Page 46 Electrical connection Proline Promass O 500 PROFIBUS DP Connecting the sensor connection housing via terminals For the device version with the order code for "Sensor connection housing": Option B "Stainless" 10 (0.4) 8 mm 22 mm 24 mm A0029613 1. Release the securing screw of the housing cover.
  • Page 47 Proline Promass O 500 PROFIBUS DP Electrical connection Connecting the sensor connection housing via the connector For the device version with the order code for "Sensor connection housing": Option C "Ultra-compact hygienic, stainless" A0029615 1. Connect the protective ground. 2. Connect the connector.
  • Page 48 Electrical connection Proline Promass O 500 PROFIBUS DP Connecting the connecting cable to the transmitter TX 20 TX 20 10 (0.4) 22 mm 24 mm A0029597 1. Loosen the 4 fixing screws on the housing cover. 2. Open the housing cover.
  • Page 49: Connecting The Signal Cable And The Supply Voltage Cable

    Proline Promass O 500 PROFIBUS DP Electrical connection 7.2.2 Connecting the signal cable and the supply voltage cable A0028200 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output Terminal connection for connecting cable between sensor and transmitter Terminal connection for signal transmission, input/output;...
  • Page 50 Electrical connection Proline Promass O 500 PROFIBUS DP WARNING Housing degree of protection may be voided due to insufficient sealing of the housing. ‣ Screw in the screw without using any lubricant. WARNING Excessive tightening torque applied to the fixing screws! Risk of damaging the plastic transmitter.
  • Page 51: Connecting The Measuring Device: Proline 500

    Proline Promass O 500 PROFIBUS DP Electrical connection Connecting the measuring device: Proline 500 NOTICE Limitation of electrical safety due to incorrect connection! ‣ Have electrical connection work carried out by appropriately trained specialists only. ‣ Observe applicable federal/national installation codes and regulations.
  • Page 52 Electrical connection Proline Promass O 500 PROFIBUS DP Connecting the sensor connection housing via terminals For the device version with the order code for "Housing": Option L "Cast, stainless" 10 (0.4) 3 mm 22 mm 24 mm A0029612 1. Loosen the securing clamp of the housing cover.
  • Page 53 Proline Promass O 500 PROFIBUS DP Electrical connection Connecting the sensor connection housing via terminals For the device version with the order code for "Housing": Option B "Stainless" 10 (0.4) 8 mm 22 mm 24 mm A0029613 1. Release the securing screw of the housing cover.
  • Page 54: Ensuring Potential Equalization

    Electrical connection Proline Promass O 500 PROFIBUS DP Connecting the connecting cable to the transmitter 10 (0.4) 3 mm 22 mm 24 mm A0029592 1. Loosen the securing clamp of the connection compartment cover. 2. Unscrew the connection compartment cover.
  • Page 55: Special Connection Instructions

    Proline Promass O 500 PROFIBUS DP Electrical connection Special connection instructions 7.5.1 Connection examples PROFIBUS DP A0028765  16 Connection example for PROFIBUS DP, non-hazardous area and Zone 2/Div. 2 Control system (e.g. PLC) Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable...
  • Page 56 Electrical connection Proline Promass O 500 PROFIBUS DP 4...20 mA A0028759  18 Connection example for 4-20 mA current output (passive) Automation system with current input (e.g. PLC) Active barrier for power supply (e.g. RN221N) Analog display unit: observe maximum load...
  • Page 57 Proline Promass O 500 PROFIBUS DP Electrical connection Relay output A0028760  21 Connection example for relay output (passive) Automation system with relay input (e.g. PLC) Power supply Transmitter: Observe input values →  255 Current input A0028915  22...
  • Page 58: Hardware Settings

    Electrical connection Proline Promass O 500 PROFIBUS DP Hardware settings 7.6.1 Setting the device address The address must always be configured for a PROFIBUS DP/PA device. The valid address range is between 1 and 126. In a PROFIBUS DP/PA network, each address can only be assigned once.
  • Page 59: Enabling The Terminating Resistor

    Proline Promass O 500 PROFIBUS DP Electrical connection Proline 500 transmitter Hardware addressing not used A0029637 Set the desired device address using the DIP switches in the connection compartment. A0029633 To switch addressing from software addressing to hardware addressing: set the DIP switch to On.
  • Page 60: Activating The Default Ip Address

    Electrical connection Proline Promass O 500 PROFIBUS DP Proline 500 – digital transmitter A0029675 1. Open the housing cover. 2. Remove the display module. 3. Fold open the terminal cover. 4. Set DIP switch No. 3 to ON. Proline 500 transmitter ‣...
  • Page 61 Proline Promass O 500 PROFIBUS DP Electrical connection A0034500 1. Loosen the 4 fixing screws on the housing cover. 2. Open the housing cover. 3. Fold open the terminal cover. 4. Set DIP switch No. 2 on the I/O electronics module from OFF → ON.
  • Page 62: Ensuring The Degree Of Protection

    Electrical connection Proline Promass O 500 PROFIBUS DP Ensuring the degree of protection The measuring device fulfills all the requirements for the IP66/67 degree of protection, Type 4X enclosure. To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following steps after the electrical connection: 1.
  • Page 63: Operation Options

    Proline Promass O 500 PROFIBUS DP Operation options Operation options Overview of operation options A0034513 Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370...
  • Page 64: Structure And Function Of The Operating Menu

    Operation options Proline Promass O 500 PROFIBUS DP Structure and function of the operating menu 8.2.1 Structure of the operating menu For an overview of the operating menu for experts: "Description of Device Parameters" document supplied with the device→  278...
  • Page 65: Operating Philosophy

    Proline Promass O 500 PROFIBUS DP Operation options 8.2.2 Operating philosophy The individual parts of the operating menu are assigned to certain user roles (operator, maintenance etc.). Each user role contains typical tasks within the device lifecycle. For custody transfer, once the device has been put into circulation or sealed, its operation is restricted.
  • Page 66: Access To The Operating Menu Via The Local Display

    Operation options Proline Promass O 500 PROFIBUS DP Menu/parameter User role and tasks Content/meaning Expert function-oriented Tasks that require detailed Contains all the parameters of the device and makes it possible to access these knowledge of the function of the parameters directly using an access code.
  • Page 67 Proline Promass O 500 PROFIBUS DP Operation options Display area In the display area, each measured value is prefaced by certain symbol types for further description: Measured variable Measurement channel Diagnostic behavior number ↓ ↓ ↓ Example Appears only if a diagnostics event is present for this measured variable.
  • Page 68: Navigation View

    Operation options Proline Promass O 500 PROFIBUS DP 8.3.2 Navigation view In the submenu In the wizard /../Operation 0091-1 /../Select medium Access stat.disp Select medium Operator Liquid Locking status Display A0013993-EN A0013995-EN Navigation view Navigation path to current position Status area Display area for navigation Operating elements →...
  • Page 69 Proline Promass O 500 PROFIBUS DP Operation options Display area Menus Symbol Meaning Operation Appears: • In the menu next to the "Operation" selection • At the left in the navigation path in the Operation menu Setup Appears: • In the menu next to the "Setup" selection •...
  • Page 70: Editing View

    Operation options Proline Promass O 500 PROFIBUS DP 8.3.3 Editing view Numeric editor +0.000 Xx – A0034250  25 For entering values in parameters (e.g. limit values) Entry display area Input screen Confirm, delete or reject entry Operating elements Text editor...
  • Page 71 Proline Promass O 500 PROFIBUS DP Operation options Operating key(s) Meaning Enter key • Press the key briefly: confirm your selection. • Press the key for 2 s: confirm the entry. Escape key combination (press keys simultaneously) Close the editing view without accepting the changes.
  • Page 72: Operating Elements

    Operation options Proline Promass O 500 PROFIBUS DP 8.3.4 Operating elements Operating key(s) Meaning Minus key In a menu, submenu Moves the selection bar upwards in a picklist. With a Wizard Confirms the parameter value and goes to the previous parameter.
  • Page 73 Proline Promass O 500 PROFIBUS DP Operation options Calling up and closing the context menu The user is in the operational display. 1. Press the  and  keys for longer than 3 seconds.  The context menu opens. XXXXXXXXXX Setup Conf.backup...
  • Page 74: Navigating And Selecting From List

    Operation options Proline Promass O 500 PROFIBUS DP 8.3.6 Navigating and selecting from list Different operating elements are used to navigate through the operating menu. The navigation path is displayed on the left in the header. Icons are displayed in front of the individual menus.
  • Page 75: Calling Up Help Text

    Proline Promass O 500 PROFIBUS DP Operation options The direct access code consists of a 5-digit number (at maximum) and the channel number, which identifies the channel of a process variable: e.g. 00914-2. In the navigation view, this appears on the right-hand side in the header of the selected parameter.
  • Page 76: User Roles And Related Access Authorization

    Operation options Proline Promass O 500 PROFIBUS DP Ent. access code Invalid or out of range input value Min:0 Max:9999 A0014049-EN For a description of the editing view - consisting of the text editor and numeric editor - with symbols →  70, for a description of the operating elements →  72 8.3.10...
  • Page 77: Enabling And Disabling The Keypad

    Proline Promass O 500 PROFIBUS DP Operation options 2. Enter the access code.  The -symbol in front of the parameters disappears; all previously write- protected parameters are now re-enabled. 8.3.12 Enabling and disabling the keypad lock The keypad lock makes it possible to block access to the entire operating menu via local operation.
  • Page 78: Prerequisites

    Operation options Proline Promass O 500 PROFIBUS DP 8.4.2 Prerequisites Computer hardware Hardware Interface CDI-RJ45 WLAN Interface The computer must have an RJ45 The operating unit must have a interface. WLAN interface. Connection Standard Ethernet cable with RJ45 Connection via Wireless LAN.
  • Page 79: Establishing A Connection

    Proline Promass O 500 PROFIBUS DP Operation options Measuring device: Via CDI-RJ45 service interface Device CDI-RJ45 service interface Measuring device The measuring device has an RJ45 interface. Web server Web server must be enabled; factory setting: ON  For information on enabling the Web server →  83...
  • Page 80 Operation options Proline Promass O 500 PROFIBUS DP IP address 192.168.1.XXX; for XXX all numerical sequences except: 0, 212 and 255 → e.g. 192.168.1.213 Subnet mask 255.255.255.0 Default gateway 192.168.1.212 or leave cells empty Via WLAN interface Configuring the Internet protocol of the mobile terminal NOTICE If the WLAN connection is lost during the configuration, settings made may be lost.
  • Page 81: Logging On

    Proline Promass O 500 PROFIBUS DP Operation options 2. Enter the IP address of the Web server in the address line of the Web browser: 192.168.1.212  The login page appears. 2 3 4 A0029417 Picture of device Device name...
  • Page 82: User Interface

    Operation options Proline Promass O 500 PROFIBUS DP 8.4.5 User interface A0029418 Function row Local display language Navigation area Header The following information appears in the header: • Device name • Device tag • Device status with status signal →  181 •...
  • Page 83: Disabling The Web Server

    Proline Promass O 500 PROFIBUS DP Operation options Navigation area If a function is selected in the function bar, the submenus of the function open in the navigation area. The user can now navigate through the menu structure. Working area...
  • Page 84: Access To The Operating Menu Via The Operating Tool

    Operation options Proline Promass O 500 PROFIBUS DP Access to the operating menu via the operating tool The structure of the operating menu in the operating tools is the same as for operation via the local display. 8.5.1 Connecting the operating tool Via PROFIBUS DP network This communication interface is available in device versions with PROFIBUS DP.
  • Page 85 Proline Promass O 500 PROFIBUS DP Operation options Proline 500 – digital transmitter open press A0029163  29 Connection via service interface (CDI-RJ45) Computer with Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for accessing the integrated device Web server or with "FieldCare", "DeviceCare" operating tool with COM DTM "CDI Communication TCP/IP"...
  • Page 86 Operation options Proline Promass O 500 PROFIBUS DP A0034569 Transmitter with integrated WLAN antenna Transmitter with external WLAN antenna LED lit constantly: WLAN reception is enabled on measuring device LED flashing: WLAN connection established between operating unit and measuring device Computer with WLAN interface and Web browser (e.g.
  • Page 87: Fieldcare

    FieldCare Function scope FDT-based plant asset management tool from Endress+Hauser. It can configure all smart field devices in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
  • Page 88: Devicecare

    Display area for current measured values Edit toolbar with additional functions such as save/restore, event list and create documentation Navigation area with operating menu structure Working area Range of action Status area 8.5.3 DeviceCare Function scope Tool to connect and configure Endress+Hauser field devices. Endress+Hauser...
  • Page 89 Proline Promass O 500 PROFIBUS DP Operation options The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. For details, see Innovation Brochure IN01047S Source for device description files See information →...
  • Page 90: System Integration

    System integration Proline Promass O 500 PROFIBUS DP System integration Overview of device description files 9.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the Operating instructions • On the transmitter nameplate • Firmware version Diagnostics →...
  • Page 91: Manufacturer-Specific Gsd

    Proline Promass O 500 PROFIBUS DP System integration 9.2.1 Manufacturer-specific GSD This GSD guarantees the unrestricted functionality of the measuring device. Device-specific process parameters and functions are therefore available. Manufacturer-specific GSD ID number File name PROFIBUS DP 0x156F EH3x156F.gsd The fact that the manufacturer-specific GSD should be used is specified in the Ident number selector parameter by selecting the Manufacturer option.
  • Page 92: Manual Setting

    System integration Proline Promass O 500 PROFIBUS DP 9.3.2 Manual setting The manual setting is made in the Ident number selector parameter via the Promass 83 (0x1529) option. Afterwards the Promass 500 PROFIBUS DP makes the same input and output data and measured value status information →...
  • Page 93: Using The Control_Block Module

    Proline Promass O 500 PROFIBUS DP System integration 9.4.1 Using the CONTROL_BLOCK module in the previous model If the CONTROL_BLOCK module is used in the previous model, the control variables are processed further if relevant functionalities can be assigned for the Promass 500.
  • Page 94: Cyclic Data Transmission

    System integration Proline Promass O 500 PROFIBUS DP Cyclic data transmission Cyclic data transmission when using the device master file (GSD). 9.5.1 Block model The block model shows which input and output data the measuring device makes available for cyclic data exchange. Cyclic data exchange takes place with a PROFIBUS master (Class 1), e.g.
  • Page 95 Proline Promass O 500 PROFIBUS DP System integration AI module (Analog Input) Transmit an input variable from the measuring device to the PROFIBUS master (Class 1). The selected input variable, along with the status, is cyclically transmitted to the PROFIBUS Master (Class 1) via the AI module. The input variable is depicted in the first four bytes in the form of a floating point number as per the IEEE 754 standard.
  • Page 96 System integration Proline Promass O 500 PROFIBUS DP Input variable Alternative GSV flow NSV flow Alternative NSV flow S&W volume flow Water cut percentage Oil density Water density Oil mass flow Water mass flow Oil volume flow Water volume flow...
  • Page 97 Proline Promass O 500 PROFIBUS DP System integration Selection: totalizer value Input variable Mass flow Volume flow Corrected volume flow Target fluid mass flow Carrier mass flow Only available with the "Concentration" application package Factory setting Function block Factory setting: TOTAL...
  • Page 98 System integration Proline Promass O 500 PROFIBUS DP Input data of TOTAL Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Measured value: floating point number (IEEE 754) Status SETTOT_MODETOT_TOTAL module The module combination consists of the SETTOT, MODETOT and TOTAL functions: •...
  • Page 99 Proline Promass O 500 PROFIBUS DP System integration Assigned compensation values A compensation value is permanently assigned to the individual Analog Output blocks. Function block Compensation value AO 1 External pressure AO 2 External temperature AO 3 External reference density AO 4 External S&W percentage...
  • Page 100 System integration Proline Promass O 500 PROFIBUS DP Data structure Input data of Discrete Input Byte 1 Byte 2 Discrete Status DO module (Discrete Output) Transmit discrete output values from the PROFIBUS master (Class 1) to the measuring device. Discrete output values are used by the PROFIBUS master (Class 1) to enable and disable device functions.
  • Page 101: Address Shifting Configuration

    Proline Promass O 500 PROFIBUS DP System integration Assignment of medium type: function block DO 7 HFCS55 HFCS90 Original wort % mass / % volume Coef Set No. 1 Coef Set No. 2 Coef Set No. 3 Hydrochloric acid Sulfuric acid...
  • Page 102 System integration Proline Promass O 500 PROFIBUS DP The configuration area contains the indexes 190 to 221 with which up to 16 parameters can be managed. Two indexes are used per parameter: • The first index is for the slot value of the parameter •...
  • Page 103: Configuring Address Shifting

    1001 : °C 1002 : °F 1000 : K 1003 : °R For more information on the "slot/index table", please contact the Endress+Hauser Sales Center. 9.6.4 Accessing data via PROFIBUS DP The PROFIBUS master uses the indexes 230 to 245 in slot 0 to access the address shifting data area.
  • Page 104 System integration Proline Promass O 500 PROFIBUS DP parameter via address shifting, the master can read out the current volume flow measured value in slot 0 and index 230. The data type (integer/float) and data access (read/write) depend on the parameter entered in the configuration area.
  • Page 105: Commissioning

    Proline Promass O 500 PROFIBUS DP Commissioning Commissioning 10.1 Function check Before commissioning the measuring device: ‣ Make sure that the post-installation and post-connection checks have been performed. • "Post-installation check" checklist→  35 • "Post-connection check" checklist →  62 10.2...
  • Page 106: Configuring The Measuring Device

    Commissioning Proline Promass O 500 PROFIBUS DP X X X X X X X 20.50 Main menu 0104-1 Display language English Operation Setup Display language 0104-1 English à Deutsch Español Français Display language 0104-1 à English Deutsch Español Français Hauptmenü...
  • Page 107: Defining The Tag Name

    Proline Promass O 500 PROFIBUS DP Commissioning Navigation "Setup" menu  Setup Device tag →  108 ‣ System units →  108 ‣ Medium selection →  111 ‣ Communication →  112 ‣ Analog inputs →  114 ‣...
  • Page 108: Setting The System Units

    Commissioning Proline Promass O 500 PROFIBUS DP XXXXXXXXX A0029422  33 Header of the operational display with tag name Tag name Enter the tag name in the "FieldCare" operating tool →  88 Navigation "Setup" menu → Device tag Parameter overview with brief description...
  • Page 109 Proline Promass O 500 PROFIBUS DP Commissioning Temperature unit →  110 Pressure unit →  110 Parameter overview with brief description Parameter Description Selection Factory setting Mass flow unit Select mass flow unit. Unit choose list Country-specific: • kg/h Result •...
  • Page 110 Commissioning Proline Promass O 500 PROFIBUS DP Parameter Description Selection Factory setting Temperature unit Select temperature unit. Unit choose list Country-specific: • °C Result • °F The selected unit applies for: • Electronic temperature parameter (6053) • Maximum value parameter (6051) •...
  • Page 111: Selecting And Setting The Medium

    Proline Promass O 500 PROFIBUS DP Commissioning 10.6.3 Selecting and setting the medium The Select medium wizard submenu contains parameters that must be configured in order to select and set the medium. Navigation "Setup" menu → Select medium ‣ Medium selection Select medium →...
  • Page 112: Configuring Communication

    Commissioning Proline Promass O 500 PROFIBUS DP Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Select medium – Select medium type. • Liquid Liquid • Gas Select gas type The Gas option is selected in Select measured gas type.
  • Page 113 Proline Promass O 500 PROFIBUS DP Commissioning Navigation "Setup" menu → Communication ‣ Communication →  113 Device address Parameter overview with brief description Parameter Description User entry Factory setting Device address Enter device address. 0 to 126 Endress+Hauser...
  • Page 114: Configuring The Analog Inputs

    Commissioning Proline Promass O 500 PROFIBUS DP 10.6.5 Configuring the analog inputs The Analog inputs submenu guides the user systematically to the individual Analog input 1 to n submenu. From here you get to the parameters of the individual analog input.
  • Page 115 Proline Promass O 500 PROFIBUS DP Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Channel – Select the process variable. • Mass flow Mass flow • Volume flow • Corrected volume flow • Density • Reference density •...
  • Page 116: Displaying The I/O Configuration

    Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting entry Fail safe type – Select the failure mode. • Fail-safe value • Fallback value • Off Fail-safe value In Fail safe type parameter, Specify the values to be output...
  • Page 117: Configuring The Current Input

    Proline Promass O 500 PROFIBUS DP Commissioning Parameter Description User interface / Selection /  Factory setting User entry Apply I/O configuration Apply parameterization of the freely • No configurable I/O module. • Yes Alteration code Enter the code in order to change the I/O Positive integer configuration.
  • Page 118: Configuring The Status Input

    Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Current span – Select current range for • 4...20 mA Country-specific: process value output and • 4...20 mA NAMUR • 4...20 mA NAMUR upper/lower level for alarm •...
  • Page 119: Configuring The Current Output

    Proline Promass O 500 PROFIBUS DP Commissioning Parameter Description User interface / Selection /  Factory setting User entry Active level Define input signal level at which the • High High assigned function is triggered. • Low Response time status input Define the minimum amount of time the...
  • Page 120 Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description User interface /  Factory setting Selection / User entry Assign current output 1 to n – Select process variable for • Off Mass flow current output. • Mass flow • Volume flow • Corrected volume flow •...
  • Page 121 Proline Promass O 500 PROFIBUS DP Commissioning Parameter Prerequisite Description User interface /  Factory setting Selection / User entry 0/4 mA value One of the following options is Enter 4 mA value. Signed floating-point Country-specific: selected in the Current span number • 0 kg/h parameter (→...
  • Page 122: 10.6.10 Configuring The Pulse/Frequency/Switch Output

    Commissioning Proline Promass O 500 PROFIBUS DP 10.6.10 Configuring the pulse/frequency/switch output The Pulse/frequency/switch output wizard guides you systematically through all the parameters that can be set for configuring the selected output type. Navigation "Setup" menu → Advanced setup → Pulse/frequency/switch output ‣...
  • Page 123 Proline Promass O 500 PROFIBUS DP Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse Pulse frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 124 Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting interface / User entry Failure mode The Pulse option is selected in Define output behavior in • Actual value No pulses the Operating mode alarm condition. • No pulses parameter (→...
  • Page 125 Proline Promass O 500 PROFIBUS DP Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting interface / User entry Operating mode – Define the output as a pulse, • Pulse Pulse frequency or switch output. • Frequency • Switch Terminal number –...
  • Page 126 Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting interface / User entry Assign frequency output The Frequency option is Select process variable for • Off selected in the Operating frequency output. • Mass flow mode parameter (→  122) •...
  • Page 127 Proline Promass O 500 PROFIBUS DP Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Maximum frequency value The Frequency option is Enter maximum frequency. 0.0 to 10 000.0 Hz 10 000.0 Hz selected in the Operating mode parameter (→  122)
  • Page 128 Commissioning Proline Promass O 500 PROFIBUS DP Configuring the switch output Navigation "Setup" menu → Pulse/frequency/switch output ‣ Pulse/frequency/switch output 1 to n Operating mode →  128 Terminal number →  128 Signal mode →  128 →  129...
  • Page 129 Proline Promass O 500 PROFIBUS DP Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch output function The Switch option is selected Select function for switch • Off in the Operating mode output. • On parameter. • Diagnostic behavior •...
  • Page 130: 10.6.11 Configuring The Relay Output

    Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting interface / User entry Assign flow direction check • The Switch option is Select process variable for flow • Off Mass flow selected in the Operating direction monitoring. • Volume flow mode parameter.
  • Page 131 Proline Promass O 500 PROFIBUS DP Commissioning Assign flow direction check →  131 Assign limit →  132 Assign diagnostic behavior →  132 Assign status →  132 →  132 Switch-off value Switch-on value →  132 Failure mode →...
  • Page 132 Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting interface / User entry Assign limit The Limit option is selected in Select process variable for limit • Mass flow Mass flow the Relay output function function. • Volume flow parameter parameter.
  • Page 133: 10.6.12 Configuring The Local Display

    Proline Promass O 500 PROFIBUS DP Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Switch-on delay In the Relay output function Define delay for the switch-on 0.0 to 100.0 s 0.0 s parameter, the Limit option is of status output.
  • Page 134 Commissioning Proline Promass O 500 PROFIBUS DP Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 135 Proline Promass O 500 PROFIBUS DP Commissioning Parameter Prerequisite Description Selection / User Factory setting entry Value 1 display A local display is provided. Select the measured value that • Mass flow Mass flow is shown on the local display. • Volume flow •...
  • Page 136 Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting entry • Pressure 0% bargraph value 1 A local display is provided. Enter 0% value for bar graph Signed floating-point Country-specific: display. number • 0 kg/h • 0 lb/min 100% bargraph value 1 A local display is provided.
  • Page 137: 10.6.13 Configuring The Low Flow Cut Off

    Proline Promass O 500 PROFIBUS DP Commissioning 10.6.13 Configuring the low flow cut off The Low flow cut off wizard systematically guides the user through all the parameters that must be set to configure low flow cut off. Navigation "Setup" menu → Low flow cut off ‣...
  • Page 138: Configuring The Partial Filled Pipe

    Commissioning Proline Promass O 500 PROFIBUS DP 10.6.14 Configuring the partial filled pipe detection The Partial filled pipe detection wizard guides you systematically through all parameters that have to be set for configuring the monitoring of the pipe filling. Navigation "Setup"...
  • Page 139: Advanced Settings

    Proline Promass O 500 PROFIBUS DP Commissioning 10.7 Advanced settings The Advanced setup submenu together with its submenus contains parameters for specific settings. Navigation to the "Advanced setup" submenu X X X X X X X 20.50 0104-1 Main menu...
  • Page 140: Calculated Values

    Commissioning Proline Promass O 500 PROFIBUS DP ‣ WLAN settings →  148 ‣ Concentration ‣ Heartbeat setup ‣ Configuration backup →  149 ‣ →  151 Administration 10.7.1 Calculated values The Calculated values submenu contains parameters for calculating the corrected volume flow.
  • Page 141: Carrying Out A Sensor Adjustment

    Proline Promass O 500 PROFIBUS DP Commissioning Parameter Prerequisite Description Selection / User Factory setting interface / User entry Fixed reference density The Fixed reference density Enter fixed value for reference Positive floating- 1 kg/Nl option is selected in the density. point number Corrected volume flow calculation parameter parameter.
  • Page 142: Configuring The Totalizer

    Commissioning Proline Promass O 500 PROFIBUS DP Navigation "Setup" menu → Advanced setup → Sensor adjustment → Zero point adjustment ‣ Zero point adjustment →  142 Zero point adjustment control →  142 Progress Parameter overview with brief description...
  • Page 143 Proline Promass O 500 PROFIBUS DP Commissioning Parameter overview with brief description Parameter Description Selection Factory setting Assign process variable Select process variable for totalizer. • Mass flow Mass flow • Volume flow • Corrected volume flow • Target mass flow •...
  • Page 144: Carrying Out Additional Display Configurations

    Commissioning Proline Promass O 500 PROFIBUS DP 10.7.4 Carrying out additional display configurations In the Display submenu you can set all the parameters associated with the configuration of the local display. Navigation "Setup" menu → Advanced setup → Display ‣...
  • Page 145 Proline Promass O 500 PROFIBUS DP Commissioning Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry Format display A local display is provided. Select how measured values • 1 value, max. size 1 value, max. size are shown on the display.
  • Page 146 Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting entry Value 1 display A local display is provided. Select the measured value that • Mass flow Mass flow is shown on the local display. • Volume flow •...
  • Page 147 Proline Promass O 500 PROFIBUS DP Commissioning Parameter Prerequisite Description Selection / User Factory setting entry • Pressure 0% bargraph value 1 A local display is provided. Enter 0% value for bar graph Signed floating-point Country-specific: display. number • 0 kg/h • 0 lb/min 100% bargraph value 1 A local display is provided.
  • Page 148: Wlan Configuration

    Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting entry Display language A local display is provided. Set display language. • English English • Deutsch (alternatively, the • Français ordered language is • Español preset in the device) •...
  • Page 149: Configuration Management

    Proline Promass O 500 PROFIBUS DP Commissioning Security type →  149 WLAN passphrase →  149 Assign SSID name →  149 SSID name →  149 →  149 Apply changes Parameter overview with brief description Parameter Prerequisite Description User entry / ...
  • Page 150 Commissioning Proline Promass O 500 PROFIBUS DP Navigation "Setup" menu → Advanced setup → Configuration backup ‣ Configuration backup →  150 Operating time →  150 Last backup Configuration management →  150 Backup state →  150 Comparison result →...
  • Page 151: Using Parameters For Device Administration

    Proline Promass O 500 PROFIBUS DP Commissioning Options Description Compare The device configuration saved in the device memory is compared with the current device configuration of the HistoROM backup. Clear backup data The backup copy of the device configuration is deleted from the memory of the device.
  • Page 152: Simulation

    Commissioning Proline Promass O 500 PROFIBUS DP Using the parameter to reset the access code Navigation "Setup" menu → Advanced setup → Administration → Reset access code ‣ Reset access code Operating time →  152 Reset access code →  152...
  • Page 153 Proline Promass O 500 PROFIBUS DP Commissioning Process variable value →  154 Status input simulation →  154 Input signal level →  154 Current input 1 to n simulation →  154 Value current input 1 to n →...
  • Page 154 Commissioning Proline Promass O 500 PROFIBUS DP Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign simulation process variable – Select a process variable for • Off the simulation process that is • Mass flow activated.
  • Page 155 Proline Promass O 500 PROFIBUS DP Commissioning Parameter Prerequisite Description Selection / User Factory setting entry / User interface Value current output 1 to n In the Current output 1 to n Enter the current value for 3.59 to 22.5 mA 3.59 mA simulation parameter, the On simulation.
  • Page 156: Protecting Settings From Unauthorized

    Commissioning Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description Selection / User Factory setting entry / User interface Diagnostic event simulation – Select a diagnostic event to • Off simulate this event. • Diagnostic event picklist (depends on the category selected) Logging interval –...
  • Page 157: Write Protection Via Write Protection

    Via Web browser, FieldCare, DeviceCare (via CDI-RJ45 service interface), fieldbus For a reset code, contact your Endress+Hauser service organization. 1. Navigate to the Reset access code parameter (→  152). 2. Enter the reset code.
  • Page 158 Commissioning Proline Promass O 500 PROFIBUS DP A0029673 1. Open the housing cover. 2. Remove the display module. 3. Fold open the terminal cover. 4. Setting the write protection (WP) switch on the main electronics module to the ON position enables hardware write protection.
  • Page 159 Proline Promass O 500 PROFIBUS DP Commissioning Proline 500 A0029630 Setting the write protection (WP) switch on the main electronics module to the ON position enables hardware write protection.  In the Locking status parameter the Hardware locked option is displayed →...
  • Page 160: Operation

    Operation Proline Promass O 500 PROFIBUS DP Operation 11.1 Reading the device locking status Device active write protection: Locking status parameter Operation → Locking status Function scope of the "Locking status" parameter Options Description None The access status displayed in the Access status parameter applies →  76. Only appears on local display.
  • Page 161: Measured Variables" Submenu

    Proline Promass O 500 PROFIBUS DP Operation 11.4.1 "Measured variables" submenu The Measured variables submenu contains all the parameters needed to display the current measured values for each process variable. Navigation "Diagnostics" menu → Measured values → Measured variables ‣...
  • Page 162: Totalizer

    Operation Proline Promass O 500 PROFIBUS DP Parameter Prerequisite Description User interface Reference density – Displays the reference density currently Signed floating-point calculated. number Dependency The unit is taken from the Reference density unit parameter (→  109). Temperature –...
  • Page 163: Input Values" Submenu

    Proline Promass O 500 PROFIBUS DP Operation Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign process variable – Select process variable for • Mass flow Mass flow totalizer. • Volume flow • Corrected volume flow •...
  • Page 164 Operation Proline Promass O 500 PROFIBUS DP Navigation "Diagnostics" menu → Measured values → Input values ‣ Input values ‣ →  164 Current input 1 to n ‣ →  164 Status input 1 to n Input values of current input The Current input 1 to n submenu contains all the parameters needed to display the current measured values for every current input.
  • Page 165: Output Values

    Proline Promass O 500 PROFIBUS DP Operation 11.4.4 Output values The Output values submenu contains all the parameters needed to display the current measured values for every output. Navigation "Diagnostics" menu → Measured values → Output values ‣ Output values ‣...
  • Page 166 Operation Proline Promass O 500 PROFIBUS DP Navigation "Diagnostics" menu → Measured values → Output values → Pulse/frequency/switch output 1 to n ‣ Pulse/frequency/switch output 1 to n Output frequency 1 to n →  166 Pulse output 1 to n →...
  • Page 167: Adapting The Measuring Device To The Process

    Proline Promass O 500 PROFIBUS DP Operation 11.5 Adapting the measuring device to the process conditions The following are available for this purpose: • Basic settings using theSetup menu (→  106) • Advanced settings using theAdvanced setup submenu (→  139) 11.6...
  • Page 168 Operation Proline Promass O 500 PROFIBUS DP Function range • A total of 1000 measured values can be stored • 4 logging channels • Adjustable logging interval for data logging • Display of the measured value trend for each logging channel in the form of a chart / ../XXXXXXXX...
  • Page 169 Proline Promass O 500 PROFIBUS DP Operation ‣ Display channel 3 ‣ Display channel 4 Endress+Hauser...
  • Page 170 Operation Proline Promass O 500 PROFIBUS DP Parameter overview with brief description Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 1 The Extended HistoROM Assign process variable to • Off application package is logging channel. • Mass flow available.
  • Page 171 Proline Promass O 500 PROFIBUS DP Operation Parameter Prerequisite Description Selection / User Factory setting entry / User interface Assign channel 2 The Extended HistoROM Assign process variable to Picklist, see Assign application package is logging channel. channel 1 parameter available. (→  170) ...
  • Page 172: Diagnostics And Troubleshooting

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting 12.1 General troubleshooting For local display Error Possible causes Solution Local display dark and no output Supply voltage does not match the Apply the correct supply voltage .
  • Page 173 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting For output signals Error Possible causes Solution Signal output outside the valid Main electronics module is Order spare part →  245. range defective. Device shows correct value on local Configuration error...
  • Page 174 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Error Possible causes Solution Web browser frozen and operation Data transfer active Wait until data transfer or current no longer possible action is finished. Connection lost 1. Check cable connection and power supply.
  • Page 175: Diagnostic Information Via Light Emitting

    Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting 12.2 Diagnostic information via light emitting diodes 12.2.1 Transmitter Proline 500 – digital Different LEDs in the transmitter provide information on the device status. A0029689 Supply voltage Device status Not used...
  • Page 176: Sensor Connection Housing

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Color Meaning Service interface (CDI), Not connected or no connection established. Ethernet Link/Activity Yellow Connected and connection established. Flashing yellow Service interface active. Proline 500 Different LEDs in the transmitter provide information on the device status.
  • Page 177 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Communication Communication Alarm Power Power A0029699 Communication Device status Supply voltage Color Meaning Communication White Communication active Device status Error Flashing red Warning Supply voltage Green Supply voltage is ok Supply voltage is off or too low...
  • Page 178: Diagnostic Information On Local Display

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP 12.3 Diagnostic information on local display 12.3.1 Diagnostic message Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display.
  • Page 179 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic behavior Symbol Meaning Alarm • Measurement is interrupted. • Signal outputs and totalizers assume the defined alarm condition. • A diagnostic message is generated. Warning Measurement is resumed. The signal outputs and totalizers are not affected. A diagnostic message is generated.
  • Page 180: Calling Up Remedial Measures

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP 12.3.2 Calling up remedial measures X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage...
  • Page 181: Calling Up Remedy Information

    Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting A0031056 Status area with status signal Diagnostic information→  179 Remedy information with Service ID In addition, diagnostic events which have occurred can be shown in the Diagnostics menu: • Via parameter •...
  • Page 182: Calling Up Remedy Information

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Xxxxxx/…/…/ Device name: Xxxxxxx Mass flow: kg/h 12.34 Device tag: Xxxxxxx Volume flow: 12.34 m /h ³ Status signal: Function check (C) Xxxxxx Diagnostics 1: C485 Simu... Remedy information: Deactivate... Failure (F)
  • Page 183: Adapting The Diagnostic Information

    Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting The user is in the Diagnostics menu. 1. Call up the desired parameter. 2. On the right in the working area, mouse over the parameter.  A tool tip with remedy information for the diagnostic event appears.
  • Page 184 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Coding Measuring value (hex) Byte 5 Quality Quality Substatus Limits A0032228-EN  36 Structure of the coding byte The content of the coding byte depends on the configured failsafe mode in the particular function block.
  • Page 185 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information pertaining to the electronics: diagnostic number 200 to 399 Diagnostic number 200 to 301, 303 to 399 Measured value status (fixed assignment) Diagnostic behavior Device diagnosis Quality Coding Category...
  • Page 186: Overview Of Diagnostic Information

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP 12.7 Overview of diagnostic information The amount of diagnostic information and the number of measured variables affected increase if the measuring device has one or more application packages. In the case of some items of diagnostic information, the diagnostic behavior can be changed.
  • Page 187 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 046 Sensor limit exceeded 1. Inspect sensor 2. Check process condition Measured variable status [from the factory] Quality Good Quality substatus Maintenance demanded Coding (hex)
  • Page 188 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 062 Sensor connection faulty 1. Check or replace sensor electronic module (ISEM) 2. If available: Check connection cable between sensor and transmitter Measured variable status 3.
  • Page 189 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 082 Data storage 1. Check module connections 2. Contact service Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal...
  • Page 190 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 083 Memory content 1. Restart device 2. Restore HistoROM S-DAT backup (' D evice reset' parameter) Measured variable status 3. Replace HistoROM S-DAT Quality Quality substatus...
  • Page 191 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 140 Sensor signal asymmetrical 1. Check or replace sensor electronic module (ISEM) 2. If available: Check connection cable between sensor and transmitter Measured variable status [from the factory] 3.
  • Page 192 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 144 Measuring error too high 1. Check or change sensor 2. Check process conditions Measured variable status [from the factory] Quality Quality substatus Maintenance alarm...
  • Page 193: Diagnostic Of Electronic

    Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting 12.7.2 Diagnostic of electronic Diagnostic information Remedy instructions Short text 201 Device failure 1. Restart device 2. Contact service Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 194 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 242 Software incompatible 1. Check software 2. Flash or change main electronics module Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 195 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 252 Modules incompatible 1. Check electronic modules 2. Check if correct modules are available (e.g. NEx, Ex) Measured variable status 3. Replace electronic modules Quality...
  • Page 196 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 262 Sensor electronic connection faulty 1. Check or replace connection cable between sensor electronic module (ISEM) and main electronics Measured variable status 2. Check or replace ISEM or main electronics...
  • Page 197 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 270 Main electronic failure Change main electronic module Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal Diagnostic behavior...
  • Page 198 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 271 Main electronic failure 1. Restart device 2. Change main electronic module Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 199 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 272 Main electronic failure 1. Restart device 2. Contact service Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal...
  • Page 200 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 273 Main electronic failure Change electronic Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 201 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 276 I/O module 1 to n faulty 1. Restart device 2. Change I/O module Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 202 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 302 Device verification active Device verification active, please wait. Measured variable status Quality Good Quality substatus Function check Coding (hex) 0xBC to 0xBF Status signal...
  • Page 203 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 311 Electronic failure 1. Do not reset device 2. Contact service Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal...
  • Page 204 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 332 Writing in HistoROM backup failed Replace user interface board Ex d/XP: replace transmitter Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 205 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 372 Sensor electronic (ISEM) faulty 1. Restart device 2. Check if failure recurs Measured variable status 3. Replace sensor electronic module (ISEM) Quality Quality substatus...
  • Page 206 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 373 Sensor electronic (ISEM) faulty 1. Transfer data or reset device 2. Contact service Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 207 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 375 I/O- 1 to n communication failed 1. Restart device 2. Check if failure recurs Measured variable status 3. Replace module rack inclusive electronic modules...
  • Page 208 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 382 Data storage 1. Insert T-DAT 2. Replace T-DAT Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal Diagnostic behavior...
  • Page 209 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 383 Memory content 1. Restart device 2. Delete T-DAT via ' R eset device' parameter Measured variable status 3. Replace T-DAT Quality Quality substatus Maintenance alarm...
  • Page 210: Diagnostic Of Configuration

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 387 HistoROM backup failed Contact service organization Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal Diagnostic behavior Alarm Influenced measured variables •...
  • Page 211 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 331 Firmware update failed 1. Update firmware of device 2. Restart device Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 212 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 410 Data transfer 1. Check connection 2. Retry data transfer Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal...
  • Page 213 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 412 Processing download Download active, please wait Measured variable status Quality Uncertain Quality substatus Initial value Coding (hex) 0x4C to 0x4F Status signal Diagnostic behavior...
  • Page 214 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 437 Configuration incompatible 1. Restart device 2. Contact service Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal Diagnostic behavior...
  • Page 215 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 438 Dataset 1. Check data set file 2. Check device configuration Measured variable status 3. Up- and download new configuration Quality Uncertain Quality substatus Maintenance demanded...
  • Page 216 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 442 Frequency output 1 to n 1. Check process 2. Check frequency output settings Measured variable status [from the factory] Quality Good Quality substatus Function check...
  • Page 217 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 453 Flow override Deactivate flow override Measured variable status Quality Good Quality substatus Function check Coding (hex) 0xBC to 0xBF Status signal Diagnostic behavior Warning Influenced measured variables •...
  • Page 218 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 482 FB not Auto/Cas Set Block in AUTO mode Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior...
  • Page 219 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 485 Measured variable simulation Deactivate simulation Measured variable status Quality Good Quality substatus Function check Coding (hex) 0xBC to 0xBF Status signal Diagnostic behavior Warning Influenced measured variables •...
  • Page 220 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 491 Current output 1 to n simulation Deactivate simulation Measured variable status Quality Good Quality substatus Function check Coding (hex) 0xBC to 0xBF Status signal...
  • Page 221 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 494 Switch output simulation 1 to n Deactivate simulation switch output Measured variable status Quality Good Quality substatus Function check Coding (hex) 0xBC to 0xBF...
  • Page 222 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 497 Simulation block output Deactivate simulation Measured variable status Quality Good Quality substatus Coding (hex) 0x80 to 0x83 Status signal Diagnostic behavior Warning Influenced measured variables –...
  • Page 223 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 529 Concentration settings faulty 1. Check concentration settings 2. Check input values e.g. pressure, temperature Measured variable status Quality Quality substatus Function check Coding (hex)
  • Page 224: Diagnostic Of Process

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP 12.7.4 Diagnostic of process Diagnostic information Remedy instructions Short text 803 Current loop 1. Check wiring 2. Change I/O module Measured variable status Quality Quality substatus Process related Coding (hex) 0x28 to 0x2B...
  • Page 225 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 831 Sensor temperature too low Increase ambient temp. around the sensor housing Measured variable status [from the factory] Quality Uncertain Quality substatus Process related Coding (hex)
  • Page 226 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 832 Electronic temperature too high Reduce ambient temperature Measured variable status [from the factory] Quality Quality substatus Process related Coding (hex) 0x28 to 0x2B Status signal...
  • Page 227 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 833 Electronic temperature too low Increase ambient temperature Measured variable status [from the factory] Quality Quality substatus Process related Coding (hex) 0x28 to 0x2B Status signal...
  • Page 228 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 834 Process temperature too high Reduce process temperature Measured variable status [from the factory] Quality Uncertain Quality substatus Process related Coding (hex) 0x78 to 0x7B...
  • Page 229 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 835 Process temperature too low Increase process temperature Measured variable status [from the factory] Quality Uncertain Quality substatus Process related Coding (hex) 0x78 to 0x7B...
  • Page 230 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 842 Process limit Low flow cut off active! 1. Check low flow cut off configuration Measured variable status [from the factory] Quality Uncertain Quality substatus...
  • Page 231 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 882 Input signal 1. Check input configuration 2. Check external device or process conditions Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 232 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 910 Tubes not oscillating 1. Check electronic 2. Inspect sensor Measured variable status Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27 Status signal...
  • Page 233 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 912 Medium inhomogeneous 1. Check process cond. 2. Increase system pressure Measured variable status [from the factory] Quality Uncertain Quality substatus Process related Coding (hex)
  • Page 234 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 913 Medium unsuitable 1. Check process conditions 2. Check electronic modules or sensor Measured variable status [from the factory] Quality Uncertain Quality substatus Process related...
  • Page 235 Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting Diagnostic information Remedy instructions Short text 942 API density out of specification 1. Check process density with selected API commodity group 2. Check API related parameters Measured variable status Quality Quality substatus...
  • Page 236 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Diagnostic information Remedy instructions Short text 944 Monitoring failed Check process conditions for Heartbeat Monitoring Measured variable status [from the factory] Quality Quality substatus Maintenance alarm Coding (hex) 0x24 to 0x27...
  • Page 237: Pending Diagnostic Events

    Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting 12.8 Pending diagnostic events The Diagnostics menu allows the user to view the current diagnostic event and the previous diagnostic event separately. To call up the measures to rectify a diagnostic event: •...
  • Page 238: 12.10 Event Logbook

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP / ../Diagnose list Diagnostics F273 Main electronic Diagnostics 2 Diagnostics 3 A0014006-EN  37 Taking the example of the local display To call up the measures to rectify a diagnostic event: •...
  • Page 239: 12.10.2 Filtering The Event Logbook

    Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting 12.10.2 Filtering the event logbook Using the Filter options parameter you can define which category of event message is displayed in the Events list submenu. Navigation path Diagnostics → Event logbook → Filter options Filter categories •...
  • Page 240: 12.11 Resetting The Measuring Device

    Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Info number Info name I1451 Monitoring on I1457 Measured error verification failed I1459 I/O module verification failed I1460 HBSI verification failed I1461 Sensor verification failed I1462 Sensor electronic module verific. failed...
  • Page 241: 12.11.1 Function Scope Of The "Device Reset" Parameter

    Proline Promass O 500 PROFIBUS DP Diagnostics and troubleshooting 12.11.1 Function scope of the "Device reset" parameter Options Description Cancel No action is executed and the user exits the parameter. To delivery settings Every parameter for which a customer-specific default setting was ordered is reset to this customer-specific value.
  • Page 242 Diagnostics and troubleshooting Proline Promass O 500 PROFIBUS DP Parameter overview with brief description Parameter Description User interface Factory setting Device tag Shows name of measuring point. Max. 32 characters, such as Promass_500 letters, numbers or special characters (e.g. @, %, /).
  • Page 243: 12.13 Firmware History

    "Manufacturer' s information" document. The manufacturer' s information is available: • In the Download Area of the Endress+Hauser web site: www.endress.com → Downloads • Specify the following details: –...
  • Page 244: Maintenance

    13.2 Measuring and test equipment Endress+Hauser offers a wide variety of measuring and test equipment, such as W@M or device tests. Your Endress+Hauser Sales Center can provide detailed information on the services.
  • Page 245: Repairs

    • The measuring devices have a modular design. • Spare parts are grouped into logical kits with the associated Installation Instructions. • Repairs are carried out by Endress+Hauser Service or by appropriately trained customers. • Certified devices can only be converted to other certified devices by Endress+Hauser Service or at the factory.
  • Page 246: Disposal

    Repairs Proline Promass O 500 PROFIBUS DP 14.5 Disposal 14.5.1 Removing the measuring device 1. Switch off the device. WARNING Danger to persons from process conditions. ‣ Beware of hazardous process conditions such as pressure in the measuring device, high temperatures or aggressive fluids.
  • Page 247: Accessories

    Various accessories, which can be ordered with the device or subsequently from Endress +Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 248: For The Sensor

    Is used to stabilize the temperature of the fluids in the sensor. Water, water vapor and other non-corrosive liquids are permitted for use as fluids.  If using oil as a heating medium, please consult with Endress+Hauser.  Special Documentation SD02159D 15.2...
  • Page 249: System Components

     Operating Instructions BA00027S and BA00059S DeviceCare Tool to connect and configure Endress+Hauser field devices.  Innovation brochure IN01047S 15.3 System components...
  • Page 250: Technical Data

    Technical data Proline Promass O 500 PROFIBUS DP Technical data 16.1 Application The measuring device is suitable for flow measurement of liquids and gases only. Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
  • Page 251: Input

    Proline Promass O 500 PROFIBUS DP Technical data 16.3 Input Measured variable Direct measured variables • Mass flow • Density • Temperature Calculated measured variables • Volume flow • Corrected volume flow • Reference density Measuring range Measuring ranges for liquids Measuring range full scale values ...
  • Page 252 • Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure measuring device for absolute pressure, e.g. Cerabar M or Cerabar S) •...
  • Page 253 Proline Promass O 500 PROFIBUS DP Technical data Input signal level • Low signal: DC –3 to +5 V • High signal: DC 12 to 30 V Assignable functions • Off • Reset the individual totalizers separately • Reset all totalizers •...
  • Page 254: Output

    Technical data Proline Promass O 500 PROFIBUS DP 16.4 Output Output signal PROFIBUS DP Signal encoding NRZ code Data transfer 9.6 kBaud…12 MBaud Current output 0/4 to 20 mA Current output 0/4 to 20 mA Maximum output values 22.5 mA...
  • Page 255 Proline Promass O 500 PROFIBUS DP Technical data Maximum pulse rate 10 000 Impulse/s Pulse value Adjustable Assignable measured • Mass flow variables • Volume flow • Corrected volume flow Frequency output Maximum input values DC 30 V, 250 mA (passive) Maximum output current 22.5 mA (active)
  • Page 256 Technical data Proline Promass O 500 PROFIBUS DP Switching behavior Can be set to: • NO (normally open), factory setting • NC (normally closed) Maximum switching • DC 30 V, 0.1 A capacity (passive) • AC 30 V, 0.5 A Assignable functions •...
  • Page 257 Proline Promass O 500 PROFIBUS DP Technical data 0 to 20 mA Failure mode Choose from: • Maximum alarm: 22 mA • Freely definable value between: 0 to 20.5 mA Pulse/frequency/switch output Pulse output Failure mode Choose from: • Actual value •...
  • Page 258: Power Supply

    Technical data Proline Promass O 500 PROFIBUS DP Light emitting diodes (LED) Status information Status indicated by various light emitting diodes The following information is displayed depending on the device version: • Supply voltage active • Data transmission active • Device alarm/error has occurred ...
  • Page 259: Performance Characteristics

    Proline Promass O 500 PROFIBUS DP Technical data Current consumption Transmitter • Max. 400 mA (24 V) • Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz) Power supply failure Depending on the device version, the configuration is retained in the device memoryor in the pluggable data memory (HistoROM DAT).
  • Page 260 Technical data Proline Promass O 500 PROFIBUS DP Density (liquids) Under Standard density Wide-range 2) 3) reference operating conditions calibration Density specification [g/cm³] [g/cm³] [g/cm³] ±0.0005 ±0.01 ±0.001 Valid over the entire temperature and density range Valid range for special density calibration: 0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F) Order code for "Application package", option EE "Special density"...
  • Page 261 Proline Promass O 500 PROFIBUS DP Technical data Current output Accuracy ±5 µA Pulse/frequency output o.r. = of reading Accuracy Max. ±50 ppm o.r. (over the entire ambient temperature range) Repeatability o.r. = of reading; 1 g/cm = 1 kg/l; T = medium temperature Base repeatability Design fundamentals →...
  • Page 262 Technical data Proline Promass O 500 PROFIBUS DP Wide-range density specification (special density calibration) If the process temperature is outside the valid range (→  259) the measured error is ±0.00005 g/cm /°C (±0.000025 g/cm /°F) [kg/m ] [°C] 40 80 120 160 200 240 280 320 360 400 [°F]...
  • Page 263: Installation

    Proline Promass O 500 PROFIBUS DP Technical data Calculation of the maximum repeatability as a function of the flow rate Flow rate Maximum repeatability in % o.r. ⋅ ½ ZeroPoint ⋅ ³ BaseRepeat A0021340 A0021335 ⋅ ½ ZeroPoint ZeroPoint ⋅...
  • Page 264: Process

    Technical data Proline Promass O 500 PROFIBUS DP External WLAN antenna IP67 Vibration resistance Device version: Order code for "Sensor connection housing", option: A "Aluminum, coated", B "Stainless" • Vibration, sinusoidal according to IEC 60068-2-6 – 2 to 8.4 Hz, 7.5 mm peak –...
  • Page 265 Proline Promass O 500 PROFIBUS DP Technical data Dependency of ambient temperature on medium temperature A0031121  39 Exemplary representation, values in the table below. Ambient temperature Medium temperature Maximum permitted medium temperature T at T = 60 °C (140 °F); higher medium temperatures T...
  • Page 266 Technical data Proline Promass O 500 PROFIBUS DP If the sensor is to be purged with gas (gas detection), it should be equipped with purge connections. Do not open the purge connections unless the containment can be filled immediately with a dry, inert gas. Use only low pressure to purge. Maximum pressure: 5 bar (72.5 psi).
  • Page 267: 16.10 Mechanical Construction

    Proline Promass O 500 PROFIBUS DP Technical data • The minimum recommended full scale value is approx. 1/20 of the maximum full scale value • In most applications, 20 to 50 % of the maximum full scale value can be considered ideal •...
  • Page 268 Technical data Proline Promass O 500 PROFIBUS DP Weight [lbs] [in] 1 261 Materials Transmitter housing Housing of Proline 500 – digital transmitter Order code for "Transmitter housing": • Option A "Aluminum coated": aluminum, AlSi10Mg, coated • Option D "Polycarbonate": polycarbonate Housing of Proline 500 transmitter Order code for "Transmitter housing":...
  • Page 269 Proline Promass O 500 PROFIBUS DP Technical data Cable entries and adapters Material Cable gland M20 × 1.5 Plastic • Adapter for cable entry with internal thread G ½" Nickel-plated brass • Adapter for cable entry with internal thread NPT ½"...
  • Page 270: 16.11 Operability

    Technical data Proline Promass O 500 PROFIBUS DP External WLAN antenna • Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickel-plated brass • Adapter: Stainless steel and nickel-plated brass • Cable: Polyethylene • Plug: Nickel-plated brass • Angle bracket: Stainless steel Process connections Fixed flange connections: –...
  • Page 271 Proline Promass O 500 PROFIBUS DP Technical data A0028232  41 Operation with touch control Proline 500 – digital Proline 500 Display elements • 4-line, illuminated, graphic display • White background lighting; switches to red in event of device errors •...
  • Page 272 Technical data Proline Promass O 500 PROFIBUS DP Supported operating Operating unit Interface Additional information tools Web browser Notebook, PC or tablet • CDI-RJ45 service Special Documentation for with Web browser interface device →  279 • WLAN interface DeviceCare SFE100 Notebook, PC or tablet •...
  • Page 273 Proline Promass O 500 PROFIBUS DP Technical data HistoROM The measuring device features HistoROM data management. HistoROM data management data management comprises both the storage and import/export of key device and process data, making operation and servicing far more reliable, secure and efficient.
  • Page 274: 16.12 Certificates And Approvals

    The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications...
  • Page 275 • With the identification PED/G1/x (x = category) on the sensor nameplate, Directive Endress+Hauser confirms conformity with the "Essential Safety Requirements" specified in Appendix I of the Pressure Equipment Directive 2014/68/EU. • Devices not bearing this marking (PED) are designed and manufactured according to good engineering practice.
  • Page 276: 16.13 Application Packages

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 277: 16.14 Accessories

    Proline Promass O 500 PROFIBUS DP Technical data Heartbeat Technology Package Description Heartbeat Verification Heartbeat Verification +Monitoring Meets the requirement for traceable verification to DIN ISO 9001:2008 Chapter 7.6 a) "Control of monitoring and measuring equipment". • Functional testing in the installed state without interrupting the process.
  • Page 278: 16.15 Supplementary Documentation

    For an overview of the scope of the associated Technical Documentation, refer to the following: • W@M Device Viewer (www.endress.com/deviceviewer): Enter the serial number from nameplate • Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2D matrix code (QR code) on the nameplate Standard documentation Brief Operating Instructions...
  • Page 279 Proline Promass O 500 PROFIBUS DP Technical data Special Documentation Contents Documentation code Information on the Pressure Equipment Directive SD01614D Radio approvals for WLAN interface for A309/A310 display module SD01793D Web server SD02232D Heartbeat Technology SD02203D Concentration measurement SD02213D Petroleum...
  • Page 280: Index

    Index Proline Promass O 500 PROFIBUS DP Index Connection tools ......36 Context menu About this document .
  • Page 281 Proline Promass O 500 PROFIBUS DP Index Display Filtering the event logbook ....239 see Onsite display Firmware Display area Release date ......90 For operational display .
  • Page 282 Index Proline Promass O 500 PROFIBUS DP Transmitter ......18 Navigation path (navigation view) ....68 Main electronics module .
  • Page 283 Proline Promass O 500 PROFIBUS DP Index Pulse/frequency/switch output (Wizard) 122, Replacement 124, Device components ..... . . 245 Pulse/frequency/switch output 1 to n (Submenu) 165 Requirements for personnel .
  • Page 284 Index Proline Promass O 500 PROFIBUS DP In the navigation view ..... 68 System design Status signals ......178, 181 Measuring system .
  • Page 285 Proline Promass O 500 PROFIBUS DP Index W@M ......244, 245 W@M Device Viewer .
  • Page 286 www.addresses.endress.com...

Table of Contents