Endress+Hauser Proline Promass O 300 Technical Information
Endress+Hauser Proline Promass O 300 Technical Information

Endress+Hauser Proline Promass O 300 Technical Information

Coriolis flowmeter
Table of Contents

Advertisement

TI01275D/06/EN/04.18
71415041
2018-11-01
The robust high‐pressure flowmeter with a compact, easily accessible
transmitter
Application
• Measuring principle operates independently of physical fluid
properties such as viscosity or density
• For premium accuracy at highest process pressures, fully
suitable for offshore conditions
Device properties
• Measuring tube in 25Cr Duplex, 1.4410 (UNS S32750)
• Process pressure up to PN 250 (Class 1500)
• Nominal diameter: DN 80 to 250 (3 to 10")
• Compact dual-compartment housing with up to 3 I/Os
• Backlit display with touch control and WLAN access
• Remote display available
Products
Technical Information
Proline Promass O 300
Coriolis flowmeter
Solutions
Your benefits
• Maximum safety – highest resistance to
stress corrosion cracking
• Fewer process measuring points – multivariable
measurement (flow, density, temperature)
• Space‐saving installation – no inlet/outlet run needs
• Full access to process and diagnostic information –
numerous, freely combinable I/Os and fieldbuses
• Reduced complexity and variety – freely configurable I/O
functionality
• Integrated verification – Heartbeat Technology
Services

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Promass O 300

  • Page 1 Products Solutions Services TI01275D/06/EN/04.18 71415041 2018-11-01 Technical Information Proline Promass O 300 Coriolis flowmeter The robust high‐pressure flowmeter with a compact, easily accessible transmitter Application Your benefits • Measuring principle operates independently of physical fluid • Maximum safety – highest resistance to...
  • Page 2: Table Of Contents

    Proline Promass O 300 Table of contents About this document ......Climate class .......
  • Page 3 Proline Promass O 300 Ordering information ..... . . 88 Application packages ..... . . 88 Diagnostics functions .
  • Page 4: About This Document

    Proline Promass O 300 About this document Symbols used Electrical symbols Symbol Meaning Direct current Alternating current Direct current and alternating current Ground connection A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system.
  • Page 5: Function And System Design

    Proline Promass O 300 Symbols in graphics Symbol Meaning 1, 2, 3, ... Item numbers , … Series of steps A, B, C, ... Views A-A, B-B, C-C, ... Sections Hazardous area Safe area (non-hazardous area) Flow direction Function and system design Measuring principle The measuring principle is based on the controlled generation of Coriolis forces.
  • Page 6: Measuring System

    Proline Promass O 300 Density measurement The measuring tube is continuously excited at its resonance frequency. A change in the mass and thus the density of the oscillating system (comprising measuring tube and fluid) results in a corresponding, automatic adjustment in the oscillation frequency. Resonance frequency is thus a function of medium density.
  • Page 7: Equipment Architecture

    Proline Promass O 300 Equipment architecture A0027512  1 Possibilities for integrating measuring devices into a system Control system (e.g. PLC) Connecting cable (0/4 to 20 mA HART etc.) Fieldbus Segment coupler Non-hazardous area Hazardous area: Zone 2; Class I, Division 2 Hazardous area: Zone 1;...
  • Page 8 Proline Promass O 300 Protecting access via hardware write protection Write access to the device parameters via the local display, Web browser or operating tool (e.g. FieldCare, DeviceCare) can be disabled via a write protection switch (DIP switch on the motherboard).
  • Page 9 Proline Promass O 300 The device can communicate with OPC UA clients using the "OPC UA Server" application package. The OPC UA server integrated in the device can be accessed via the WLAN access point using the WLAN interface - which can be ordered as an optional extra - or the service interface (CDI- RJ45) via Ethernet network.
  • Page 10: Input

    Proline Promass O 300 Input Measured variable Direct measured variables • Mass flow • Density • Temperature Calculated measured variables • Volume flow • Corrected volume flow • Reference density Measuring range Measuring range for liquids Measuring range full scale values ...
  • Page 11: Operable Flow Range

    • Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure measuring device for absolute pressure, e.g. Cerabar M or Cerabar S) • Medium temperature to increase accuracy (e.g. iTEMP) •...
  • Page 12 Proline Promass O 300 Status input Maximum input values • DC –3 to 30 V • If status input is active (ON): R >3 kΩ Response time Adjustable: 5 to 200 ms Input signal level • Low signal: DC –3 to +5 V •...
  • Page 13: Output

    Proline Promass O 300 Output Output and input variants Depending on the option selected for output/input 1, different options are available for the other outputs and inputs. Only one option can be selected for each output/input 1 to 3. The table must be read vertically (↓).
  • Page 14: Output Signal

    Proline Promass O 300 Output signal HART current output Current output 4 to 20 mA HART Current span Can be set to: 4 to 20 mA (active/passive)  Ex-i, passive Open-circuit voltage DC 28.8 V (active) Maximum input voltage DC 30 V (passive) Load 250 to 700 Ω...
  • Page 15 Proline Promass O 300 Current consumption 10 mA Permitted supply voltage 9 to 32 V Bus connection With integrated reverse polarity protection Modbus RS485 Physical interface RS485 in accordance with EIA/TIA-485 standard Terminating resistor Integrated, can be activated via DIP switches...
  • Page 16 Proline Promass O 300 Open-circuit voltage DC 28.8 V (active) Pulse width Adjustable: 0.05 to 2 000 ms Maximum pulse rate 10 000 Impulse/s Pulse value Adjustable Assignable measured • Mass flow variables • Volume flow • Corrected volume flow...
  • Page 17 Proline Promass O 300 Double pulse output Function Double pulse Version Open collector Can be set to: • Active • Passive Maximum input values DC 30 V, 250 mA (passive) Open-circuit voltage DC 28.8 V (active) Voltage drop For 22.5 mA: ≤ DC 2 V...
  • Page 18: Signal On Alarm

    Proline Promass O 300 The technical values correspond to those of the inputs and outputs described in this section. Signal on alarm Depending on the interface, failure information is displayed as follows: HART current output Device diagnostics Device condition can be read out via HART Command 48...
  • Page 19 Proline Promass O 300 0 to 20 mA Failure mode Choose from: • Maximum alarm: 22 mA • Freely definable value between: 0 to 20.5 mA Pulse/frequency/switch output Pulse output Failure mode Choose from: • Actual value • No pulses...
  • Page 20: Ex Connection Data

    Proline Promass O 300 Web server Plain text display With information on cause and remedial measures Light emitting diodes (LED) Status information Status indicated by various light emitting diodes The following information is displayed depending on the device version: • Supply voltage active •...
  • Page 21: Low Flow Cut Off

    Proline Promass O 300 Intrinsically safe values Order code for Output type Intrinsically safe values "Output; input 1" "Output; input 1" 26 (+) 27 (–) Option CA Current output = 30 V 4 to 20 mA HART Ex i = 100 mA = 1.25 W...
  • Page 22 Proline Promass O 300 PROFIBUS PA Manufacturer ID 0x11 Ident number 0x156D Profile version 3.02 Device description files (GSD, Information and files under: DTM, DD) • www.endress.com • www.profibus.org Supported functions • Identification & Maintenance Simplest device identification on the part of the control system and nameplate •...
  • Page 23 Proline Promass O 300 Compatibility with If the device is replaced, the measuring device Promass 300 supports the earlier model compatibility of the cyclic data with previous models. It is not necessary to adjust the engineering parameters of the PROFIBUS network with the Promass 300 GSD file.
  • Page 24 Proline Promass O 300 Conformity class Conformance Class B Netload Class Netload Class II Baud rates Automatic 100 Mbit/s with full-duplex detection Cycle times From 8 ms Polarity Auto-polarity for automatic correction of crossed TxD and RxD pairs Media Redundancy Protocol...
  • Page 25 Proline Promass O 300 ITK Test Campaign Number Information: • www.endress.com • www.fieldbus.org Link Master capability (LAS) Choice of "Link Master" and "Basic Device" Factory setting: Basic Device Node address Factory setting: 247 (0xF7) Supported functions The following methods are supported: •...
  • Page 26: Terminal Assignment

    Proline Promass O 300 Supported baud rate • 1 200 BAUD • 2 400 BAUD • 4 800 BAUD • 9 600 BAUD • 19 200 BAUD • 38 400 BAUD • 57 600 BAUD • 115 200 BAUD Data transfer mode •...
  • Page 27: Device Plugs Available

    Proline Promass O 300 Modbus RS485 Supply voltage Input/output 1 Input/output 2 Input/output 3 1 (+) 2 (–) 26 (B) 27 (A) 24 (+) 25 (–) 22 (+) 23 (–) The terminal assignment depends on the specific device version ordered →  13.
  • Page 28: Pin Assignment, Device Plug

    Proline Promass O 300 Order code for "Input; output 1", option RA "PROFINET" Order code for Cable entry/connection →  31 "Electrical connection" L, N, P, U Connector M12 × 1 – 1) 2) 1) 2) 1) 2) 1) 2) Connector M12 ×...
  • Page 29 Proline Promass O 300 PROFINET Assignment TD + RD + TD – RD – Coding Plug/socket A0032047 Socket  Recommended plug: • Binder, series 763, part no. 99 3729 810 04 • Phoenix, part no. 1543223 SACC-M12MSD-4Q • When using the device in a hazardous location, use a suitably certified plug.
  • Page 30: Power Consumption

    Proline Promass O 300 Power consumption Transmitter Max. 10 W (active power) switch-on current Max. 36 A (as per NAMUR Recommendation NE21) Current consumption Transmitter • Max. 400 mA (24 V) • Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz) Power supply failure •...
  • Page 31: Electrical Connection

    Proline Promass O 300 Electrical connection Connecting the transmitter • Terminal assignment→  26 • Device plugs available→  27 A0026781 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output or terminal for network connection via service interface (CDI-RJ45);...
  • Page 32 Proline Promass O 300 Connecting the remote display and operating module DKX001 The remote display and operating module DKX001 is available as an optional extra →  89. • The measuring device is always supplied with a dummy cover when the remote display and operating module DKX001 is ordered directly with the measuring device.
  • Page 33 Proline Promass O 300 4...20 mA A0028762  3 Connection example for 4 to 20 mA HART current output (passive) Automation system with current input (e.g. PLC) Power supply Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable specifications →...
  • Page 34 Proline Promass O 300 PROFIBUS PA A0028768  5 Connection example for PROFIBUS PA Control system (e.g. PLC) PROFIBUS PA segment coupler Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable...
  • Page 35 Proline Promass O 300 EtherNet/IP A0028767  7 Connection example for EtherNet/IP Control system (e.g. PLC) Ethernet switch Observe cable specifications Device plug Transmitter EtherNet/IP: DLR (Device Level Ring) A0027544 Control system (e.g. PLC) Ethernet switch Observe cable specifications →  40...
  • Page 36 Proline Promass O 300 PROFINET: MRP (Media Redundancy Protocol) A0027544 Control system (e.g. PLC) Ethernet switch Observe cable specifications →  40 Connecting cable between the two transmitters Transmitter FOUNDATION Fieldbus A0028768  9 Connection example for FOUNDATION Fieldbus Control system (e.g. PLC) Power Conditioner (FOUNDATION Fieldbus) Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements;...
  • Page 37 Proline Promass O 300 Modbus RS485 A0028765  10 Connection example for Modbus RS485, non-hazardous area and Zone 2; Class I, Division 2 Control system (e.g. PLC) Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable...
  • Page 38 Proline Promass O 300 Pulse/frequency output 12345 A0028761  13 Connection example for pulse/frequency output (passive) Automation system with pulse/frequency input (e.g. PLC) Power supply Transmitter: Observe input values →  15 Switch output A0028760  14 Connection example for switch output (passive) Automation system with switch input (e.g.
  • Page 39 Proline Promass O 300 A0029279  16 Connection example for double pulse output (passive) Automation system with double pulse input (e.g. PLC) Power supply Transmitter: Observe input values →  17 Double pulse output Double pulse output (slave), phase-shifted Relay output A0028760 ...
  • Page 40: Potential Equalization

    Proline Promass O 300 Status input A0028764  19 Connection example for status input Automation system with status output (e.g. PLC) Power supply Transmitter Potential equalization Requirements No special measures for potential equalization are required. Please consider the following to ensure correct measurement: •...
  • Page 41 Proline Promass O 300 Cable type Characteristic impedance 135 to 165 Ω at a measuring frequency of 3 to 20 MHz Cable capacitance < 30 pF/m Wire cross-section > 0.34 mm (22 AWG) Cable type Twisted pairs Loop resistance ≤110 Ω/km Signal damping Max.
  • Page 42 Proline Promass O 300 Pulse/frequency/switch output Standard installation cable is sufficient. Double pulse output Standard installation cable is sufficient. Relay output Standard installation cable is sufficient. Current input 0/4 to 20 mA Standard installation cable is sufficient. Status input Standard installation cable is sufficient.
  • Page 43: Performance Characteristics

    Proline Promass O 300 Performance characteristics Reference operating • Error limits based on ISO 11631 conditions • Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi) • Specifications as per calibration protocol •...
  • Page 44: Repeatability

    Proline Promass O 300 SI units 1:10 1:20 1:50 1:100 1:500 [mm] [kg/h] [kg/h] [kg/h] [kg/h] [kg/h] [kg/h] 180 000 18 000 9 000 3 600 1 800 350 000 35 000 17 500 7 000 3 500 800 000...
  • Page 45: Influence Of Ambient Temperature

    Proline Promass O 300 Influence of ambient Current output temperature Temperature coefficient Max. 1 μA/°C Pulse/frequency output Temperature coefficient No additional effect. Included in accuracy. Influence of medium Mass flow and volume flow temperature o.f.s. = of full scale value When there is a difference between the temperature for zero point adjustment and the process temperature, the additional measured error of the sensor is typically ±0.0002 % o.f.s./°C (±0.0001 %...
  • Page 46: Design Fundamentals

    Proline Promass O 300 [% o.r./bar] [% o.r./psi] [mm] [in] –0.002 –0.0001 –0.002 –0.0001 o.r. = of reading, o.f.s. = of full scale value Design fundamentals BaseAccu = base accuracy in % o.r., BaseRepeat = base repeatability in % o.r.
  • Page 47: Mounting Location

    Proline Promass O 300 Mounting location A0028772 To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid the following mounting locations in the pipe: • Highest point of a pipeline. • Directly upstream of a free pipe outlet in a down pipe.
  • Page 48: Inlet And Outlet Runs

    Proline Promass O 300 Orientation Recommendation Vertical orientation A0015591 Horizontal orientation, transmitter at Exceptions: →  21,  48 A0015589 Horizontal orientation, transmitter at bottom Exceptions: →  21,  48 A0015590 Horizontal orientation, transmitter at side A0015592 This orientation is recommended to ensure self-draining.
  • Page 49 Proline Promass O 300 WARNING Danger from medium escaping! Medium escaping under pressure can cause injury or material damage. ‣ Take precautions to prevent danger to persons and damage if the rupture disk is actuated. ‣ Observe information on the rupture disk sticker.
  • Page 50: Environment

    Dependency of ambient temperature on medium temperature→  51 ‣ If operating outdoors: Avoid direct sunlight, particularly in warm climatic regions. You can order a weather protection cover from Endress+Hauser. →  89. Storage temperature –50 to +80 °C (–58 to +176 °F) Climate class...
  • Page 51: Interior Cleaning

    Proline Promass O 300 Interior cleaning • Cleaning in place (CIP) • Sterilization in place (SIP) Options • Oil- and grease-free version for wetted parts, without declaration Order code for "Service", option HA • Oil- and grease-free version for wetted parts as per IEC/TR 60877-2.0 and BOC 50000810-4, with declaration Order code for "Service", option HB...
  • Page 52: Pressure-Temperature Ratings

    Proline Promass O 300 Pressure-temperature The following pressure/temperature diagrams apply to all pressure-bearing parts of the device and not just the process connection. The diagrams show the maximum permissible medium pressure ratings depending on the specific medium temperature. Flange connection according to EN 1092-1 (DIN 2501)
  • Page 53: Sensor Housing

    Proline Promass O 300 Flange connection according to ASME B16.5 [psi] [bar] 4000 Cl. 1500 3600 3200 2800 Cl. 900 2400 2000 Cl. 00 1600 1200 Cl. 00 [°C] [°F] A0036636-EN  24 With flange material stainless steel, 1.4410/F53 25Cr Duplex (Super Duplex)
  • Page 54: Rupture Disk

    Proline Promass O 300 If there is a need to drain the leaking medium into a discharge device, the sensor should be fitted with a rupture disk. Connect the discharge to the additional threaded connection →  62. If the sensor is to be purged with gas (gas detection), it should be equipped with purge connections.
  • Page 55: System Pressure

    Proline Promass O 300 System pressure It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas. This is prevented by means of a sufficiently high system pressure. For this reason, the following mounting locations are recommended: •...
  • Page 56: Custody Transfer Measurement

    After putting the device into circulation or after sealing the device, operation is only possible to a limited extent. Detailed ordering information is available from your local Endress+Hauser sales center for national approvals, which are based on the OIML certificates, for applications with liquids other than water or gases.
  • Page 57: Mechanical Construction

    Proline Promass O 300 Mechanical construction Dimensions in SI units Compact version A0033786 Order code for "Housing", option A "Aluminum, coated" [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] 40.5 51.2 68.9 99.1 Depending on the cable gland used: values up to + 30 mm...
  • Page 58 Proline Promass O 300 [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] 68.9 1 009 99.1 Depending on the cable gland used: values up to + 30 mm Dependent on respective process connection Endress+Hauser...
  • Page 59 Proline Promass O 300 Flange connections Fixed flange EN 1092-1, ASME B16.5, JIS B2220 A0023178 Length tolerance for dimension L in mm: • DN ≤ 100: +1.5 / –2.0 • DN ≥ 125: +3.5 Flange according to EN 1092-1 Form B1 (DIN 2501): PN16 Stainless steel, 1.4410/F53 25Cr Duplex (Super Duplex)
  • Page 60 Proline Promass O 300 Flange according to EN 1092-1 Form B2 (DIN 2501): PN160 Stainless steel, 1.4410/F53 25Cr Duplex (Super Duplex) Order code for "Process connection", option DAD [mm] [mm] [mm] [mm] [mm] [mm] [mm] 8 × Ø26 80.9 8 × Ø30 104.3...
  • Page 61 Proline Promass O 300 Flange according to ASME B16.5: Class 300 Schedule 40 Stainless steel, 1.4410/F53 25Cr Duplex (Super Duplex) Order code for "Process connection", option ABD [mm] [mm] [mm] [mm] [mm] [mm] [mm] 387.4 16 × Ø28.6 48.1 254.5 1 862 Flange according to ASME B16.5: Class 600 Schedule 40...
  • Page 62 Proline Promass O 300 RTJ flange according to ASME B16.5: Class 1500 Schedule 80 Stainless steel, 1.4410/F53 25Cr Duplex (Super Duplex) Order code for "Process connection", option AGD [mm] [mm] [mm] [mm] [mm] [mm] [mm] 203.2 8 × Ø31.8 55.7 73.7...
  • Page 63 Proline Promass O 300 [mm] [mm] [mm] Protective cover 280 (11.0) 255 (10.0) 146 (5.75) 134 (5.3) 12 (0.47) 30 (1.18) A0029553 Remote display and operating module DKX001 78 (3.07) 136 (5.35) 114 (4.49) 103 (4.06) 7.0 (0.27) A0028921  28...
  • Page 64: Dimensions In Us Units

    Proline Promass O 300 External WLAN antenna mounted on device 105 (4.1) 68 (2.7) 173 (6.8) A0028923  29 Engineering unit mm (in) External WLAN antenna mounted with cable The external WLAN antenna can be mounted separately from the transmitter if the transmission/ reception conditions at the transmitter mounting location are poor.
  • Page 65 Proline Promass O 300 Order code for "Housing", option A "Aluminum, coated" [in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in] [in] 6.65 2.68 3.98 7.87 14.45 22.32 7.87 2.32 5.55 1.59 5.47 6.65 2.68 3.98 15.08 25.08...
  • Page 66 Proline Promass O 300 Flange connections Fixed flange ASME B16.5 A0023178 Length tolerance for dimension L in inch: • DN ≤ 4": +0.06 / –0.08 • DN ≥ 5": +0.14 Flange according to ASME B16.5: Class 150 Schedule 40 Stainless steel, 1.4410/F53 25Cr Duplex (Super Duplex) Order code for "Process connection", option AAD...
  • Page 67 Proline Promass O 300 Flange according to ASME B16.5: Class 1500 Schedule 80 Stainless steel, 1.4410/F53 25Cr Duplex (Super Duplex) Order code for "Process connection", option AFD [in] [in] [in] [in] [in] [in] [in] 10.43 8 × Ø1.0 2.16 39.09 12.20...
  • Page 68 Proline Promass O 300 Accessories Purge connections/rupture disk DN 8 ( ")...150 (6") DN 250 (10") ⁄ 45° 35 (1.38) ca. 75 (approx. 2.95) SW 1" A0028914 Connection nipple for purge connections: order code for "Sensor options", option CH "Purge connection"...
  • Page 69 Proline Promass O 300 Remote display and operating module DKX001 78 (3.07) 136 (5.35) 114 (4.49) 103 (4.06) 7.0 (0.27) A0028921  31 Engineering unit mm (in) External WLAN antenna The external WLAN antenna is not suitable for use in hygienic applications.
  • Page 70: Weight

    Proline Promass O 300 70 (2.8) A0033597  33 Engineering unit mm (in) Weight All values (weight exclusive of packaging material) refer to devices with ASME B16.5 Class 900 flanges. Weight specifications including transmitter as per order code for "Housing", option A "Aluminum, coated".
  • Page 71 Proline Promass O 300 Cable entries/cable glands A0028352  34 Possible cable entries/cable glands Female thread M20 × 1.5 Cable gland M20 × 1.5 Adapter for cable entry with internal thread G ½" or NPT ½" Device plugs Order code for "Housing", option A "Aluminum, coated"...
  • Page 72: Process Connections

    Proline Promass O 300 Seals Welded process connections without internal seals Accessories Protective cover Stainless steel, 1.4404 (316L) External WLAN antenna • Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickel-plated brass • Adapter: Stainless steel and nickel-plated brass • Cable: Polyethylene •...
  • Page 73: Local Operation

    Proline Promass O 300 Local operation Via display module Two display modules are available: • Order code for "Display; operation", option F "4-line, illuminated, graphic display; touch control" • Order code for "Display; operation", option G "4-line, illuminated, graphic display; touch control + WLAN"...
  • Page 74: Remote Operation

    Proline Promass O 300 Material The housing material of the display and operating module DKX001 depends on the choice of transmitter housing material. Transmitter housing Remote display and operating module Order code for "Housing" Material Material Option A "Aluminum, coated"...
  • Page 75 Proline Promass O 300 A0028746  38 Options for remote operation via HART protocol (passive) Control system (e.g. PLC) Transmitter power supply unit, e.g. RN221N (with communication resistor) Connection for Commubox FXA195 and Field Communicator 475 Field Communicator 475 Computer with Web browser (e.g. Internet Explorer) for access to the integrated device Web server or computer with an operating tool (e.g.
  • Page 76 Proline Promass O 300 Via PROFIBUS DP network This communication interface is available in device versions with PROFIBUS DP. A0020903  40 Options for remote operation via PROFIBUS DP network Automation system Computer with PROFIBUS network card PROFIBUS DP network...
  • Page 77 Proline Promass O 300 A0029437  42 Options for remote operation via Modbus-RS485 protocol (active) Control system (e.g. PLC) Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP" or Modbus DTM...
  • Page 78 Proline Promass O 300 A0033725  44 Options for remote operation via EtherNet/IP network: ring topology Automation system, e.g. "RSLogix" (Rockwell Automation) Workstation for measuring device operation: with Custom Add-On Profile for "RSLogix 5000" (Rockwell Automation) or with Electronic Data Sheet (EDS) Computer with Web browser (e.g.
  • Page 79: Service Interface

    Proline Promass O 300 A0033719  46 Options for remote operation via PROFINET network: ring topology Automation system, e.g. Simatic S7 (Siemens) Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or computer with operating tool (e.g. FieldCare, DeviceCare, SIMATIC PDM) with COM DTM "CDI Communication TCP/IP"...
  • Page 80: Network Integration

    Proline Promass O 300 A0034570 Transmitter with integrated WLAN antenna Transmitter with external WLAN antenna LED lit constantly: WLAN reception is enabled on measuring device LED flashing: WLAN connection established between operating unit and measuring device Computer with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for accessing the integrated device Web server or with operating tool (e.g.
  • Page 81: Supported Operating Tools

    Proline Promass O 300 A0033618 Automation system, e.g. Simatic S7 (Siemens) Ethernet switch Edge Gateway Cloud Measuring device Ethernet network Measured values via inputs and outputs Optional WLAN interface The optional WLAN interface is available on the following device version: Order code for "Display;...
  • Page 82: Historom Data Management

    Proline Promass O 300 Supported operating Operating unit Interface Additional information tools FieldCare SFE500 Notebook, PC or tablet • CDI-RJ45 service →  91 with Microsoft Windows interface system • WLAN interface • Fieldbus protocol Device Xpert Field Xpert SFX...
  • Page 83 Proline Promass O 300 Additional information on the data storage concept There are different types of data storage units in which device data are stored and used by the device: Device memory T-DAT S-DAT Available data • Event logbook such as diagnostic events for example •...
  • Page 84: Certificates And Approvals

    The device meets the legal requirements of the applicable EU Directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system meets the EMC requirements of the "Australian Communications and Media...
  • Page 85: Pharmaceutical Compatibility

    Proline Promass O 300 Ex tb Category Type of protection II2D Ex tb IIIC T** °C Db Currently, the following versions for use in hazardous areas are available: IS (Ex i) and XP (Ex d) • Class I, III, III Division 1 Groups A-G •...
  • Page 86: Certification Profibus

    Directive • With the identification PED/G1/x (x = category) on the sensor nameplate, Endress+Hauser confirms conformity with the "Essential Safety Requirements" specified in Appendix I of the Pressure Equipment Directive 2014/68/EU.
  • Page 87: Other Standards And Guidelines

    Proline Promass O 300 Testing of welded connections Option Test standard Component ISO 23277 AL2x (PT) ASME ASME NORSOK Measuring tube Process connection ISO 10675-1 AL1 (RT, DR) B31.3 VIII Div.1 M-601 Appx. 4+8 VT, PT VT, RT VT, PT...
  • Page 88: Ordering Information

    The application packages can be ordered with the device or subsequently from Endress+Hauser. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 89: Concentration

    →  93. Accessories Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are available for the device. Detailed information on the order code in question is available from your local Endress+Hauser sales center or on the product page of the Endress+Hauser website: www.endress.com.
  • Page 90: Communication-Specific Accessories

    Is used to stabilize the temperature of the fluids in the sensor. Water, water vapor and other non-corrosive liquids are permitted for use as fluids.  If using oil as a heating medium, please consult with Endress+Hauser.  Special Documentation SD02159D...
  • Page 91: Service-Specific Accessories

    Accessories Description Applicator Software for selecting and sizing Endress+Hauser measuring devices: • Choice of measuring devices for industrial requirements • Calculation of all the necessary data for identifying the optimum flowmeter: e.g. nominal diameter, pressure loss, flow velocity and accuracy.
  • Page 92: System Components

    For an overview of the scope of the associated Technical Documentation, refer to the following: • W@M Device Viewer (www.endress.com/deviceviewer): Enter the serial number from nameplate • Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2D matrix code (QR code) on the nameplate Standard documentation...
  • Page 93: Device-Dependent Additional Documentation

    Proline Promass O 300 Description of Device Parameters Documentation code FOUNDATION Measuring device HART Fieldbus PROFIBUS PA PROFIBUS DP Modbus RS485 EtherNet/IP PROFINET Promass 300 GP01057D GP01094D GP01058D GP01134D GP01059D GP01114D GP01115D Device-dependent Safety instructions additional documentation Safety instructions for electrical equipment for hazardous areas.
  • Page 94: Registered Trademarks

    Proline Promass O 300 Contents Documentation code HART FOUNDATION PROFIBUS PA PROFIBUS DP Modbus RS485 PROFINET EtherNet/IP Fieldbus Web server SD01662D SD01665D SD01664D SD02226D SD01663D SD01969D SD01968D Heartbeat Technology SD01642D SD01696D SD01698D SD02202D SD01697D SD01988D SD01982 Concentration SD01644D SD01706D SD01708D...
  • Page 96 www.addresses.endress.com...

Table of Contents