Parameter
P62
the temperature above which the
condenser temperature alarm is
activated (with reference to the
condenser probe)
P63
P62 differential
P64
temperature below which the minimum
temperature alarm is activated during
post-chill storage, with relation to P7,
i.e. "P7 + P64" (with reference to the
cabinet probe) 0 = no alarm
P65
temperature above which the maximum
temperature alarm is activated during
post-chill storage, with relation to P7,
i.e. "P7
+ P65" (with reference to the cabinet
probe) 0 = no alarm
P66
temperature below which the minimum
temperature alarm is activated during
post-freezing storage, with relation to
P8, i.e. "P8 + P66" (with reference to
the cabinet probe)
0 = no alarm
P67
temperature above which the maximum
temperature alarm is activated during
post-freezing storage, with relation to
P8, i.e. "P8 + P67" (with reference to
the cabinet probe)
0 = no alarm
P68
P64, P65, P66 and P67 differential
P69
storage operation start-up temperature
alarm delay
P70
temperature alarm delay
P71
enable printing 1 = YES
P72
print interval
P73
HACCP alarm list deletion 1 = YES
P74
Baudrate 0 = 2400
1 = 4800
2 = 9600
3 = 19200
P75
Bit polarity 0 = no polarity 1 = even
2 = odd
P76
Board address
Temperature Probes Display
To check the reading of the unit's temperature sensing
probes, verify the unit is OFF.
1.
Press the NEEDLE PROBE key and the DOWN arrow key
at the same time and hold for five (5) seconds.
2.
Press the UP or DOWN arrow key to scroll through the
list.
DISPLAY
Pr1
Pr2
System Diagnosis Guide
Description
Factory
Value
70
10
0
0
0
0
2
0
0
0
5
0
2
2
1
PROBE
Cabinet Probe
Needle Probe
DISPLAY
Pr3
If the unit struggles to reach the cabinet temperature
setpoint (i.e., -40°F), obtain valuable information by
checking the evaporator probe.
Difficulty reaching this temperature may indicate that the
evaporator coil is not supplied with sufficient refrigerant.
With a cabinet setpoint of -40°F (-40°C), the evaporator
temperature is approximately -49°F (-45°C) at 0 Barg.
The controller evaporator probe reading MUST BE at least
-43.6°F (-42°C), the minimum readable temperature.
HACCP Printer
Blast Chillers can be ordered with an optional printer, or
one can be installed as a retrofit. The printer has its own
power supply connection (110VAC – 240VAC 50/60 Hz) (see
Figure 2). The printer is connected to the Main Board at the
second RS 485 connection port (see Figure 3).
Figure 2:
Printer Power Supply Connection
Figure 3: Printer Connected to Main Board
Connect the printer to the Main Board using coaxial
shielded cables.
55
PROBE
Evaporator Probe

Advertisement

Table of Contents
loading

Table of Contents