Operation
Introduction
This chapter describes how to use the test xture and the measurement error correcting
techniques for the test xture.

Operating Flow

Before performing a measurement, you have to compensate for the residual error of the test
xture by using the measurement error correcting functions. Perform the following steps when
measuring the DUT:
3
(
indicates procedures in the following pages.)
7
Described in Chapter 2.
The 16191A's electrical length is 14.0 mm (typical value).
To set the electrical length, see your analyzer's Operation Manual.
To set the xture and devices for xture compensation, see:
\How to set the SHORT condition, and the OPEN condition"
7
These procedures show how to set the test xture for the
OPEN/SHORT/LOAD compensation procedure.
For how to measure OPEN/SHORT/LOAD, see the Operation Manual
supplied with your analyzer. (You can omit LOAD compensation.)
To connect the DUT, see:
\How to connect the DUT"
7
3
Operation
3-1