Short Impedance Repeatability Check; Short Impedance Value Guideline; Short Impedance Repeatability Guideline - Agilent Technologies 16191A Operation And Service Manual

Side electrode smd test fixture
Table of Contents

Advertisement

Parameter
Ls
Ls

Short Impedance Repeatability Check

1. After the Short Impedance Check is completed, remove the Shorting Device and put it back
on the same place again.
2. Make a single sweep measurement with the same measurement conditions.
3. Read Ls value at 100 MHz and 1 GHz and check the value is within the Short Impedance
Value Guideline described in Table 4-8.
4. Calculate the di erence between the rst Ls measurement value and second Ls
measurement value at each frequency.
The guideline of the short impedance repeatability is as follows:
Parameter
Ls
Ls
Table 4-8. Short Impedance Value Guideline
Frequency
100 MHz
1 GHz
Table 4-9. Short Impedance Repeatability Guideline
Frequency
100 MHz
1 GHz
Guideline (absolute value)
6
3 nH
1 nH
6
3 nH
1 nH
Guideline (di erence)
6
45 pH
6
20 pH
Service
4-11

Advertisement

Table of Contents
loading