Residual Inductance Of The Shorting Bar; Applicable Dut Size - Agilent Technologies 16191A Operation And Service Manual

Side electrode smd test fixture
Table of Contents

Advertisement

Residual Inductance of the Shorting Bar

The usual method to compensate the test xture's residual inductance is to let SHORT = 0H.
In this method, the measurement result is the relative value of the measured impedance to the
shorting bar's impedance. The short bar's residual inductance as a result of its size and shape is
not estimated.
On the other hand, there is a de nition method to let SHORT = x H. In this method, the
measurement result is the absolute value of the device's impedance. The short bar's residual
inductance as a result of its size and shape is estimated under speci c conditions and is used as
a reference value. This method, is useful for devices with values which are close to the short
conditions of the measurement system.
The reference inductance values presented Table 1-2 and Table 1-3 were simulated as the
relative di erence to a disk-type 0 termination on either the 7 mm or the 3.5 mm connector.
The measurement of these short bars under other conditions than shown below cannot
reproduce the reference inductance values.
Table 1-2. Option 701 Shorting Device Residual Inductance (Typical)
Shorting Bar l [mm] d [mm] h [mm] O set [mm] Connector Inductance
P/N 16191-29001
P/N 16191-29002
P/N 16191-29002
P/N 16191-29002
P/N 16191-29003
P/N 16191-29004
1-4
General Information
Figure 1-1. Applicable DUT Size
1
2.4
1
1.6
2.4
2
2
2.4
1.6
2.4
2
1.6
3.2
2.4
2.4
4.5
2.4
2.4
(Typical)
0.75
3.5 mm
0.45
3.5 mm
0.25
3.5 mm
1.3
7 mm
0.9
7 mm
0
7 mm
0.2 nH
0.2 nH
0.2 nH
0.7 nH
0.6 nH
0.6 nH

Advertisement