Compensation For Fixture Residual Impedance Error - Agilent Technologies 16048A Operation And Service Manual

Test leads
Table of Contents

Advertisement

Operation

Compensation for Fixture Residual Impedance Error

Compensation for Fixture Residual Impedance Error
The 16048E has inherent stray capacitance, residual inductance, and residual
resistance that affect the accuracy of measured values. To compensate for, or
negate, these residuals to minimize measurement error, the instrument's
Open/Short compensation procedure should be performed. The procedure is given
in the instrument's operating manual.