Chapter 5 Specifications

5.1 Specifications

Output Part
Max. peak
Voltage Range
Current Range
Setting Resolution
Mode
Polarity
Setting Resolution
Measure-
Voltage Range
ment Part
Current Range
Measurement Resolution
Measurement Accuracy
Data Acqui-
Measurement cycle
sition
Measuring point
Measurement time
Pulse Width
Duty Ratio
Item
Power
(Note1)
(Note1)
(Note1)
(Note2,3,4)
(Note1)
HC UNIT CS-200 Series Instruction Manual
Specifications
25 kW (2 kA Range)
12.5 kW (1 kA Range)
6.25 kW (500A Range)
(Note2)
50 V
2 kA to 5 A (2/1/5 Step)
1 kA to 5 A (2/1/5 Step)
500 A to 5 A (2/1/5 Step)
1/±20000
PULSE
POSITIVE, NEGATIVE
± (Set value x 2% + Set Range x 0.1%)
Note1: Voltage setting is specified with no load
Note2: Limited to 40V at 2 kA setting in voltage mode
50 V to 200 mV (2/1/5 Step)
2 kA to 5 A (2/1/5 Step)
1 kA to 5 A (2/1/5 Step)
500 A to 5 A (2/1/5 Step)
1/±20000
± (Set value x 2% + Set Range x 0.1%)
Note1: Voltage measurement is measured at MAIN UNIT (using MV
SENSE terminal)
Note2: Can measure up to 125% of the measurement range
Note3: Specified within the measurement range
Note4: Specified in combination with MAIN UNIT
2 ms to 5 s
Can be set
Can be set
10 μs to 500 μs (2 kA, 1 kA)
10 μs to 1 ms (500 A to 5 A)
Note1: When the pulse width is narrow, it may not be possible to
output up to the maximum voltage / current.
≦0.1% (2 kA)
≦0.25% (1 kA)
≦0.5% (500 A, 200 A)
≦1% (100 A to 5 A)
Chapter 5 Specifications
(1/2)
(CS-220)
(CS-210)
(CS-205)
(CS-220)
(CS-210)
(CS-205)
(CS-220)
(CS-210)
(CS-205)
5-1