7.4.2

Installing the underplaten tubing

Connect tubes to nipples
Insert tubes into tube clamps
®
Hematek
3000 System – Instruction Manual – Version 1.3
You will need:
• 1 new Hematek Underplaten Tubing Set
To install the underplaten tubing, proceed as follows:
1. On the middle panel, connect one of the new tubes of the underplaten
tubing to each nipple.
1
4
1 Middle panel
2 Nipples
3 Tube clamps
4 Rear conveyor spiral
Figure 7-7 Connecting the underplaten tubing to the nipples
2. Make sure that at least 7mm (1/4 in) of each tube is connected to a
nipple.
3. Open the tube clamps.
4. Insert the tubes into the tube clamps.
5. Close the tube clamps.