Standby developer and fixer replenishment
(jog cycle)
In order to compensate for oxidation of chemicals during a protracted standby
period, an additional replenishment of developer and fixer can be initiated auto-
matically every hour ('jog cycle' function).
The default setting is AUTO: the jog cycle will be automatically switched ON or
OFF in accordance with your daily film consumption: ON if the consumption is
less than 3 m² per day, OFF if the consumption is more than 3 m² per day.
Refer to the table for recommended jog cycle settings, depending on the specific
workload.
more than 5 m² film/day
daily use
irregular use
400 ml/m²
more than 5 m² film/day
daily use
irregular use
OFF
more than 5 m² film/day
daily use
irregular use
OFF
The jog cycle is not adjustable during film processing.
If you do not comply with the recommended replenishment rates, the
optimal processing stability is no longer guaranteed and sensitometric
variations might occur.
1280C GB 20001204
Related setting: Developer replenishment rate
3 to 5 m² film/day
daily use
400 ml/m²
600 ml/m²
Standby developer and fixer replenishment - jog cycle
3 to 5 m² film/day
daily use
OFF
OFF
Related setting: Start-up replenishment function
3 to 5 m² film/day
daily use
ON
OFF
less than 3 m² film/day
irregular use
daily use
600 ml/m²
600 ml/m²
less than 3 m² film/day
irregular use
daily use
OFF
less than 3 m² film/day
irregular use
daily use
ON
OFF
irregular use
600 ml/m²
irregular use
ON
ON
irregular use
ON