Developer replenishment rate

The developer replenishment rate must be set in accordance with the amount of
film that is processed. The Classic E.O.S. allows for surface scanning of film,
taking into account not only the film length but also the film width.
The default setting is 400 ml/m².
Refer to the table for recommended developer replenishment rates, depending
on your specific workload.
more than 5 m² film/day
daily use
400 ml/m²
Related setting: Standby developer and fixer replenishment - jog cycle
more than 5 m² film/day
daily use
OFF
more than 5 m² film/day
daily use
OFF
The developer replenishment rate is not adjustable during film processing.
If you do not comply with the recommended replenishment rates, the opti-
mal processing stability is no longer guaranteed and sensitometric varia-
tions might occur.
40
Developer replenishment rate
irregular use
daily use
400 ml/m²
600 ml/m²
irregular use
daily use
OFF
OFF
Related setting: Start-up replenishment function
irregular use
daily use
ON
OFF
3 to 5 m² film/day
irregular use
600 ml/m²
3 to 5 m² film/day
irregular use
OFF
3 to 5 m² film/day
irregular use
ON
less than 3 m² film/day
daily use
irregular use
600 ml/m²
600 ml/m²
less than 3 m² film/day
daily use
irregular use
ON
ON
less than 3 m² film/day
daily use
irregular use
OFF
ON
1280C GB 20001204