Download Print this page

Installation And Startup - Siemens SIREMOBIL Installation Instructions Manual

Hide thumbs Also See for SIREMOBIL:

Advertisement

2
Safety information
WARNING
Validity of this document
This document is applicable if a key switch is to be retrofitted to the SIREMOBIL Compact,
SIREMOBIL Compact L, SIREMOBIL Iso-C, POWERMOBIL, and ARCOSKOP systems.
Documents required
System manual of the product
Parts required
Retrofit kit key switch
Tools required
Tool case
1 set of Allen keys
1 drilling machine
1 vacuum cleaner
Measuring equipment required
Protective conductor resistance meter
e. g. "Safety Tester Bender UNIMET 1000 ST", Mat. No. 51 38 727
SIREMOBIL
System Manual
Avoidance of light to severe injuries which may even lead to
death and avoidance of material damage.
The product-specific safety information given in the system man-
ual as well as the safety information contained in the ARTD,
part 2, must be observed.
After all work has been completed and all covers have been
attached, perform the protective conductor test according to
ARTD-002.731.17. The protective conductor resistance must not
exceed 0.2 Ohm.
07140895
SPR2-130.812.06
Rev. 01

Installation and Startup

Page 2 of 6
12.02
CS PS 24
1
1
1
1
1
1
1
Siemens AG