Fuse Replacement - Fluke VT Plus HF Operator's Manual

Gas flow analyzer
Hide thumbs Also See for VT Plus HF:
Table of Contents

Advertisement

VT Plus HF
Operators Manual

Fuse Replacement

The Analyzer fuses may have to be replaced. The fuse compartment is located on the side
panel of the instrument, as shown in Figure 5-1.
To replace the fuses:
1. Power off Analyzer and unplug the main power cord.
2. Using a small, flat-blade screwdriver or similar tool, carefully pry off the cover to the
3. Pull out the fuse block assembly and carefully lift the old fuses from the fuse slots.
4. Replace the fuses (two 0.5 A Slo-Blo fuses, Fluke part #46026).
To avoid possible fire, electrical shock, or personal injury, use
only the recommended fuse.
5. Replace the cover/fuse block in the fuse compartment.
6. Plug in the Analyzer.
5-4
Figure 5-1. Fuse Compartment
fuse compartment. The fuse block assembly is attached to the back of the cover.
XW Warning
fec122.eps