Fluke v3000 FC Calibration Manual

Fluke v3000 FC Calibration Manual

Wireless ac voltage
Hide thumbs Also See for v3000 FC:

Advertisement

Quick Links

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
v3000 FC
Wireless AC Voltage
Calibration Manual
July 2014
© 2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Advertisement

Table of Contents
loading

Summary of Contents for Fluke v3000 FC

  • Page 1 Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com v3000 FC Wireless AC Voltage Calibration Manual July 2014 © 2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
  • Page 2 Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction ......................1 Contact Fluke ..................... 1 Safety Information ....................2 Symbols ......................3 Specifications ..................... 4 AC Voltage ..................... 4 General ......................4 Required Equipment ..................5 Performance Tests ..................... 5 Test the Display ..................... 5 Backlight ......................
  • Page 4 FC Calibration Manual...
  • Page 5 List of Tables Table Title Page Symbols ........................3 Required Equipment ....................5 Performance Tests ....................6 Calibration Adjustment ................... 10 User-Replaceable Parts ..................12...
  • Page 6 FC Calibration Manual...
  • Page 7 List of Figures Figure Title Page All Segments of the Display ................... 5 Performance Test Connections ................7 Calibration Password Reset ................... 9 Battery Replacement ....................12...
  • Page 8 FC Calibration Manual...
  • Page 9: Introduction

    Introduction Warning Read "Safety Information" before you use the Product. This manual has the verification and calibration adjustment procedures for the v3000 FC Wireless AC Voltage module (the Product). Please see the v3000 FC Quick Reference Guide for usage information.
  • Page 10: Safety Information

    FC Calibration Manual Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.  Warning To prevent possible electrical shock, fire, or personal injury: Carefully read all instructions.
  • Page 11: Symbols

    Product Category: With reference to the equipment types in the  WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
  • Page 12: Specifications

    FC Calibration Manual Specifications Accuracy is specified as ±( [% of reading] + [number of least significant digits] ). All ranges are autoranging. Accuracy is specified from 1 % to 100 % of the range obtained by autoranging, from 18 °C to 28 °C.
  • Page 13: Required Equipment

    1. With the Product off, push and hold . 2. Push while you keep  pushed until all of the display segments are shown. See Figure 1. hby002.eps Figure 1. All Segments of the Display If segments of the display are missing, repair is necessary. See “Contact Fluke”.
  • Page 14: Backlight

    To verify that the keypad functions, turn on the Product and push each button separately. Each button push will turn on a display annunciator. If the buttons do nothing, repair is necessary. See “Contact Fluke”. AC Voltage Test Before you do the ac voltage test: 1.
  • Page 15: Before Calibration Adjustment

    Wireless AC Voltage Before Calibration Adjustment hcc07.eps Figure 2. Performance Test Connections Before Calibration Adjustment Before the Product calibration can be adjusted, you must go through the maintenance mode menu and enter your password. Maintenance Mode The Product maintenance mode can be used to set different parameters on the Product that include auto power off, backlight adjustment, and calibration.
  • Page 16: Password Entry

    FC Calibration Manual Password Entry To go to the calibration mode, push  five times until CAL is shown. You will need to enter a password to access calibration mode. To enter the password: 1. Push  and the CAL counter is shown. For example n002.
  • Page 17: Restore The Default Password

    Wireless AC Voltage Before Calibration Adjustment Restore the Default Password If the calibration password is lost, the default password (1234) can be manually restored with the subsequent steps:  Warning To prevent electric shock or personal injury, remove all input signals before you open the Product.
  • Page 18: Calibration Adjustment

    FC Calibration Manual Calibration Adjustment The Product features closed-case calibration adjustment and uses known reference sources. The Product measures the applied reference source, calculates correction factors, and stores the correction factors in nonvolatile memory. Should the Product fail any of the performance tests, do the calibration adjustment procedure.
  • Page 19: Maintenance

    Wireless AC Voltage Maintenance Maintenance Clean the Product Caution To prevent possible damage to the Product or to equipment under test, do not use abrasive cleaners. They will damage the case. To clean the Product, use a cloth with a mild cleaning solution. Battery Replacement ...
  • Page 20: User-Replaceable Parts

    FC Calibration Manual hcc001.eps Figure 4. Battery Replacement User-Replaceable Parts User-replaceable parts are shown in Table 5. Table 5. User-Replaceable Parts Fluke Part Description Number 4130305 FLK-3000-2003, DOOR, BATTERY 1881997 TPAK Meter Hanging Kit variable TEST LEADS variable LARGE ALLIGATOR CLIP BLACK, BULK...

Table of Contents