Integration - Teledyne T3PM1100 User Manual

Digital power meter
Hide thumbs Also See for T3PM1100:
Table of Contents

Advertisement

Find Quality Products Online at:
20 s
Auto ( * )
( * ) Limit of the measurement lower limit frequency
by the Timeout setting
Timeout
1 s
5 s
10 s
20 s
Auto switching among six types: 100mHz, 1 Hz, 10
Measurement range
Hz, 100 Hz, 1 kHz, 10 kHz, and 100 kHz.
Frequency filter
Select OFF or ON (cut off frequency of 500 Hz)
Requirements
Accuracy
± (0.06% of reading)

Integration

Item
Specifications
Select manual integration mode, standard integration
Mode
mode, or repetitive integration mode.
Automatically stop integration by setting a timer.
Timer
Selectable range: 0 hours 00 minutes 00 seconds to
9999 hours 59 minutes 59 seconds
±(Power accuracy (or current accuracy) + 0.1% of
Accuracy
reading) (fixed range)
Range setting
Auto range or fixed range is available for Integration
Timer accuracy
±0.02%
Start, stop and reset operations are available using
Remote control
an external remote signal. (option)
206
GlobalTestSupply
www.
T3PM1100 User Manual
0.1 Hz ≤ f ≤ 100 kHz
0.1 Hz ≤ f ≤ 100 kHz
lower limit frequency
2.0 Hz
0.5 Hz
0.2 Hz
0.1 Hz
When the input signal level is
30% or more of the
measurement range If the
crest factor is set to 3.
(60% or more if the crest
factor is set to 6 or 6A)
• Frequency filter is ON when
measuring voltage or current
of 200 Hz or less.
.com
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

Table of Contents