Appendix C:
For technical specifications on the Hettich ROTANTA 46 RSC ROBOTIC, refer
to the Hettich operating instructions included with your system.
Centrifuge Module Specifications
Table 4 lists the operating and physical specifications for the Centrifuge
Module:
Table 4:
Specification
Operating Conditions
Operating Temperature
Range
Humidity
Air Supply
Air Consumption
Dimensions
Power Requirements
Throughput
Hettich Unit Specifications
The Hettich unit can be programmed for the following:
•
Temperatures of 10 to 40C
•
Speeds up to 4500 RPM
092D0001-02 Rev. A
Specifications
Centrifuge Module Specifications
Description
The Centrifuge module is designed for safety
from voltage surges according to Installation
Category II and Pollution degree 2 (IEC EN
61010-1).
Use the Centrifuge module indoors only. Do not
operate the system at altitudes above 2000 m
(6000 ft).
15–32°C
Relative humidity from 20–80% noncondensing
60–75 psi
< 10 L/min
•
•
•
•
208–240 VAC, 50/60 Hz, 16A MAX
•
•
•
Weight 705 kg (1550 lb)
Width 1794 mm (70.61 inches)
Depth 1111 mm (43.7 inches)
Height 2121 mm (83.5 inches)
Up to 300 tubes per hour
Up to 72 tubes per cycle
Up to 18 tubes per cassette
page 91