Centrifuge Module Interface Operating Instructions
d. Wipe dry with a clean, dry cleaning cloth.
Always use a fresh dry cloth to ensure there is no dampness on the
scale.
5. Return the cassettes to the scale.
6. Move the robot so it is above the Hettich unit.
7. Close the robot enclosure door.
8. Pull out the E-stop button.
9. Bring the Centrifuge module online.

Cleaning the Interface Gate Tube Holder

BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to
Appendix A, Safety Information for recommended precautions when
working with biohazardous materials.
Cleaning Cycle
Materials Needed
To clean the Interface Gate Tube Holder, perform these steps:
1. At the LineMaster workspace, ensure the Centrifuge module is in the
shutdown state and no sample tubes remain in the Centrifuge module
2. At the Centrifuge module, push in the E-stop button.
3. To turn off power from the interface gate, push in the gate control
button on the associated divert gate.
092D0001-02 Rev. A
Weekly
Cleaning cloth
Isopropyl alcohol (70%)
Water
Maintenance
page 53