Centrifuge Module Interface Operating Instructions
3. On the back, lower-left side of the instrument, locate and turn off the
circuit breaker.
Figure 6:
1. Circuit breaker
Recovering from a Total Emergency Shutdown
CAUTION
Do not use sample tubes left on the system during an emergency
shutdown without checking sample status. After recovering from an
emergency shutdown of the Centrifuge module, check all sample tubes
removed from the Centrifuge module against the sample information in
the LineMaster to determine their current status before continuing with
testing.
When the centrifuge is ready for use after an emergency shutdown, use
the following steps to bring your system online:
1. On the back of the Centrifuge, turn on the circuit breaker.
092D0001-02 Rev. A
Centrifuge (Back View) Circuit Breaker
Troubleshooting
page 71