Troubleshooting
b. If you did not remove the sample tube, select No.
4. Repeat these steps for all missing sample tubes.
Cleaning after Sample Tube Breakage
BIOHAZARD
Wear personal protective equipment. Use universal precautions. Refer to
Appendix A, Safety Information for recommended precautions when
working with biohazardous materials.
After removing a broken tube, use the cleaning procedures in Chapter 3,
Maintenance to clean all affected areas of the Hettich unit and Centrifuge
module, as necessary:
•
Cleaning the Hettich Unit Outer Enclosure , page 59
•
Cleaning the Hettich Unit Chamber and Seals , page 63
•
Inspecting and Cleaning the Buckets and Rotor, page 60
•
Cleaning the Cassettes, page 51
•
Cleaning the Scale, page 50
•
Cleaning the Robot Gripper, page 47
•
Cleaning the Interface Gate Tube Holder, page 53
page 82
Inspect the Centrifuge module for the missing tube. If you find the
tube, you can manually remove it from the Centrifuge module.
Centrifuge Module Interface Operating Instructions
092D0001-02 Rev. A