8.1 Specifications
Item
Weight (main unit)
Consumption current
(Note 1)
The full scale (F.S.) in terms of accuracy in a voltage ranging from 1 to 5 V DC and of a current
ranging from 4 to 20 mA are 0 to 5 V DC and 0 to 20 mA, respectively.
(Note 2)
The indication accuracy when set to high-speed mode is ±0.3% (at 0 to 55°C.
(Note 3)
The indication accuracy rating of the thermocouple range varies according to the input type of the
thermocouple to be used and the measurement temperature.
(Note 4)
If highly accurate temperature data is necessary, use the temperature data about 30 minutes after the
unit was powered on.
(Note 5)
If a sudden temperature change occurs in the unit, the accuracy may temporarily not be stable. Install
the unit so as to keep it out of direct streams of wind from devices such as a control panel cooling fan.
(Note 6)
For the unit Ver. 1.0, the K2 range is -200 to +1000°C and the J2 range is -200 to +750°C.
● Also refer to the section
■
Indication accuracy rating of thermocouple range
Measurement temperature
Thermoc
(Note 1)
ouple
range
K1
-100.0
K2
-200.0
J1
-100.0
J2
-200.0
-270.0
T
-200.0
0.0
-270.0
N
-200.0
0.0
0.0
R
300.0
0.0
S
300.0
0.0
B
400.0
800.0
-270.0
E
8-4
Specifications
145 g
80mA or less
"Indication accuracy rating of thermocouple range"
Normal mode: 25 ms conversion
Ambient
temperature
25°C
to
600.0°C
±0.1%
to
1370.0°C
±0.1%
to
400.0°C
±0.1%
to
1200.0°C
±0.1%
to
-200.0°C
(Note 2)
to
0.0°C
±0.15%
to
400.0°C
±0.1%
to
-200.0°C
(Note 2)
to
0.0°C
±0.15%
to
1300.0°C
±0.1%
to
300.0°C
±0.15%
to
1760.0°C
±0.1%
to
300.0°C
±0.15%
to
1760.0°C
±0.1%
to
400.0°C
(Note 2)
to
800.0°C
±0.15%
to
1820.0°C
±0.1%
to
-200.0°C
(Note 2)
High-speed mode: 5 ms
conversion
Ambient
Ambient temperature 0
temperature 0 to
to 55°C
55°C
±0.3%
±0.30%
±0.3%
±0.30%
±0.3%
±0.30%
±0.3%
±0.30%
(Note 2)
(Note 2)
±0.3%
±0.30%
±0.3%
±0.30%
(Note 2)
(Note 2)
±0.3%
±0.30%
±0.3%
±0.30%
±0.3%
±1.00%
±0.3%
±0.30%
±0.3%
±1.00%
±0.3%
±0.30%
(Note 2)
(Note 2)
±0.3%
±1.00%
±0.3%
±0.30%
(Note 2)
(Note 2)
on the next page.
WUME-FP7TCRTD-04