Final Procedures - Siemens MAMMOMAT Nova Opdima 1000 Service Instructions Manual

Table of Contents

Advertisement

6 - 14

Final procedures

1.
Check in Database mode that the latest examination performed is available and
the images can be viewed.
2.
Check camera image quality in normal and high resolution. Use calibration
exposure data, see Appendix 2.
3.
If the image shows defects e.g. lines, spots or irregularities such as 45 degree-
mesh pattern, calibrate the camera according to CCD camera calibration and
maintenance on Page 5 - 22.
Check biopsy accuracy with the test phantom, if the accuracy is not within + 1mm
4.
in (x, y, z), calibrate the biopsy unit according to Calibration of the biopsy unit on
Page 5 - 3.
5.
Turn screen lock on/off, according to Software upgrade,
Miscellaneous on Page 5 - 18.