3 - 8
3.

Final procedure

1.
2.
3.
MAMMOMAT 3000 Modular
Service
If no discrepancies appear in 1b, the problem lies in the cassette.
If no discrepancies appear in 1c, the problem lies in the object table.
If discrepancies still appear in the same area, when comparing the original film
with the test films from 1a c, the X-ray tube window must be examined for dirt
and, if necessary, be cleaned. Should the discrepancies still remain, the fault is in
the X-ray tube.
Fit the collimator cover (1) and fasten with the four Allen screws (2) (see fig 2.1).
Check that all three anode/filter combinations can be chosen for both large/small
focus and 24
30 table (if used).
System OFF.
Reinstall the swivel-arm covers.
Register 5
SPB7-230.092.11
Rev. 01
04.98
Installation
Page 8 of 8
Siemens-Elema AB
SPS-UD
3
Solna, Sweden