Enabling The Dicom Option; Instruction For Obtaining The Systems Host Id And Hostname - Siemens MAMMOMAT Nova Opdima 1000 Service Instructions Manual

Table of Contents

Advertisement

Service mode

Enabling the DICOM option

To enable the DICOM option you need the DICOM license key that is received when
purchasing the DICOM option.
The DICOM licence is written on the invoice and the dispatch note. If not, please contact
Siemens AG Medical Solutions in Erlangen, Headquarter Support Center of Special Prod-
ucts Division (Dept. CS HSC 24) via phone (+49/ (0) 9191 18 8080.1.6.3) for further infor-
mation. This department will ask for serial No. and host ID of the workstation in question.

Instruction for obtaining the systems host ID and hostname

1.
Switch on the workstation.
2.
Log in as service user.
3.
Press the Control, Alt, Shift and ! keys at the keyboard simultaneously.
4.
Switch to default behavior?
Press OK. The screen will flash for a second.
5.
Place the mouse pointer on the screen background and press the rightmost
mouse button.
6.
Select New window.
7.
Type "hostname" and press Enter.
NOTICE
8.
Type "hostid" and press Enter.
NOTICE
9.
Type "exit" and press Enter.
10.
Press the Control, Alt, Shift and ! keys at the keyboard simultaneously.
11.
Switch to custom behavior?
Press OK. The screen will flash for a second.
12.
Done!
Siemens AG
Medical Solutions
Note the hostname (opdxxxx) shown on the screen.
Note the host id (xxxxxxxx) shown on the screen.