Service mode
CAUTION
Fine needle
Needle channel center
Fig. 9 Measuring of needle length
NOTICE

Editing needle values

1.
Select a needle from the needle setup dialog.
2.
Press Edit... .
Siemens AG
Medical Solutions
Ensure by measuring the length of the specified needle or core
gun with needle that the right values are entered.
If the stroke length of the core gun is changed, a new needle has
to be selected from the Targets dialog. The core needle length L1
shall be measured with the core needle in outer position and
mounted in the core gun.
L1
When using core gun, the system safety margin to avoid hitting
the biopsy table is 5 mm.
Register 5
SPB7-230.061.11
Rev. 02
Core gun
L2
L1
Page 7 of 30
07.03
CS PS 24
5 - 7
Core gun
with coax
5