Ccd Camera; Biopsy Controller Cable Duct - Siemens MAMMOMAT Nova Opdima 1000 Service Instructions Manual

Table of Contents

Advertisement

10 - 6

CCD camera

Use the protective ground wire tester (≤ 6V/10225A).
1.
Measure the resistance between the ground connection of the stand and
measuring points on the CCD camera according to Fig. 4. Use a sharp-pointed
tool to penetrate the finishing paint. Fill in your values in the test protocol for
protective earth measurement (see Appendix 2).
Fig. 4 Measuring points on the biopsy unit
The value of the resistance measured must not exceed 0.1 Ω.

Biopsy controller cable duct

Use the protective ground wire tester (≤ 6V/10225A).
1.
Measure the resistance between the ground connection of the stand and the
cover of the biopsy controller cable duct. Use a sharp-pointed tool to penetrate the
finishing paint. Fill in your values in the test protocol for protective earth
measurement (see Appendix 2).
The value of the resistance measured must not exceed 0.1 Ω.
Opdimar
Service
1
3
Register 5
SPB7-230.061.11
Rev. 02
Measures after service
2
Page 6 of 6
07.03
CS PS 24