TROUBLESHOOTING

If any abnormality arises during use, please check the following points
SYMPTOMS
Unit does not turn
Batteries have
on when the START/
run down.
STOP button is
pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on dis-
incorrectly.
play or the blood pres-
sure value is displayed
excessively low
Did you talk or move
(or high).
during measurement?
Shaking of the arm
with the cuff on.
E1 error symbol shown
Air circuit abnormality.
on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on display
exceeding 300 mmHg.
E3 error symbol shown
Error determining
on display
measurement data.
EP error symbol
System error.
shown on display
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair the
unit by yourself. Contact information for HoMedics Consumer Relations
Department can be found on the warranty page.
POSSIBLE
CORRECTION
CAUSES
Replace them with
four new "AA"
alkaline batteries.
Re-insert the batteries
in the correct posi-
tions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again. Keep
arm steady during
measurement.
Check cuff connection.
Measure again.
Switch the unit off,
then measure again.
Measure again.
Measure again.
35

SPECIFICATIONS

Four 1.5V DC "AA" alkaline batteries or 6V DC AC
Power Source:
adapter
Measurement Method: Oscillometric
Pressure: 40~280 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
60 memories for each User (120 total)
Memory Capacity:
1 minute after last button operation
Auto-shut-off:
Operation
Temperature: 41°F~104°F (5°C ~40°C)
Humidity: 15% ~ 93% R.H.
Environment:
Storage/Transportation
Temperature: -13°F~158°F (-25°C~70°C)
Humidity: Less Than 93% R.H.
Environment:
0.65 lbs (293 g) (without batteries)
Weight:
Standard size cuff: 9"–13" (23-33 cm)
Arm Circumference:
Large size cuff:13"–17" (33-43 cm)
5.6"(L) x 4.3"(W) x 2.5"(H)
Dimensions:
143.2mm(L) x 108.7mm(W) x 63mm(H)
(4) "AA" alkaline batteries, 6V DC AC adapter,
Accessories:
pouch, USB cable, arm cuffs with tube
Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the ingress of water
Not suitable for use in presence of flammable anes-
thetic mixture with air, oxygen or nitrous oxide.
Continuous operation with short-time loading.
Follow instructions for use.
Note: These specifications are subject to change without notice.