TROUBLESHOOTING
If
any abnormality arises
during use, please
check
the
following
points:
SYMPTOMS
POSSIBLE CAUSES
CORRECTION
Unit does not turn on when
Batteries have run down.
Replace them with two
new
"AAA"
the
START/
STOP button
is
alkaline batteries.
pushed.
Battery polarities have been
Re-insert the batteries in the correct
positioned incorrectly.
positions.
EE measurement error
symbol
Cuff has been placed
Wrap the cuff properli so that it is
shown on display or the blood
incorrectly.
positioned correctly.
easure again.
pressure value
is
displa~ed
Did you talk or move during
Keep wrist steady during
excessively low
(or
high
.
measurement?
measurement. Measure again.
Shaking of the wrist with the
cuff on.
E1 error symbol shown on
display.
Air circuit abnormality.
Check cuff placement. Measure
again.
E2 error symbol shown on
Inflation pressure exceeding
Switch the unit OFF, then measure
display.
300 mmHg.
again.
E3 error symbol shown on
display.
Can't determine blood pressure
measurement data.
Rewrap
the cuff properly, keep
steady, and measure again.