Advertisement
When using your blood pressure monitor, basic precautions should always be followed. Please read and follow all instructions and warnings before using this product. Save these instructions for future reference.
Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultatory method within the accuracy limits prescribed by the American National Standard for manual, electronic, or automated sphygmomanometers.
What is blood pressure?
Blood pressure is the pressure exerted on the artery walls while blood flows through the arteries. The pressure measured when the heart contracts and sends blood out of the heart is systolic (highest) blood pressure. The pressure measured when the heart dilates with blood flowing back into the heart is called diastolic (lowest) blood pressure.
Why measure your blood pressure?
Among today's various health problems, those associated with high blood pressure are very common. High blood pressure dangerously correlates with cardiovascular diseases. Therefore, blood pressure monitoring is important for identifying those at risk.
Why do my readings vary?
Blood pressure is a body parameter that is subject to normal variations throughout the day. A single reading that is different from your and your doctor's readings are not necessarily inaccurate. The average of several readings, taken under similar conditions, using the same arm, is preferred for accurate blood pressure readings.
Why are my readings different than those taken at my doctor's office?
Many experience a phenomenon called "White Coat Hypertension" when measured by a doctor. White Coat Hypertension refers to blood pressure that rises above its usual level when measured in a clinical setting, such as a doctor's office.
The table below contains defined levels of hypertension from the World Health Organization (WHO). The WHO has established globally accepted standards from the assessment for high blood pressure readings. Users can compare their own blood pressure readings against these defined levels to determine if they may potentially be at increased risk. This table is applicable to most adults age 18 and older.
Blood pressure tends to go up and down, even in people who normally don't have high readings. If your numbers stay above the "normal" range most of the time, you may be at increased risk and should consult your physician.
Although one can easily find where their own blood pressure readings fall on this table, this monitor comes equipped with a Risk Category Index that automatically compares each reading to the defined levels and provides a helpful cue if your reading falls into one of the stages that could potentially indicate an increased risk.
Please note that cues provided by this monitor are only intended to assist you in using this table. The table and cues are only provided for convenience to help you understand your non-invasive blood pressure reading as it relates to the WHO information. They are not a substitute for a medical examination by your physician. It is important for you to consult with your physician regularly. Your physician will tell you your normal blood pressure range as well as the point at which you may actually be considered to be at risk.
NOTE: This blood pressure monitor uses defined levels for hypertension from World Health Organization (WHO). This table provides the American Heart Association blood pressure guidelines (AHA 2017) for your reference. You can compare your readings to this chart to know where your measurement falls according to the AHA 2017 defined levels for hypertension. | Blood Pressure Category | Sistolic mmHg (upper number) | Diastolic mmHg (lower number) | |
Normal | Less than 120 | and | Less than 80 | |
Elevated | 120 – 129 | and | Less than 80 | |
High Blood Pressure (Hypertension) Stage 1 | 130 – 139 | or | 80 – 89 | |
High Blood Pressure (Hypertension) Stage 2 | 140 or higher | or | 90 or higher | |
Hypertension Crisis | Higher than 180 | and/or | Higher than 120 |
*Source: AHA 2017
Replace the batteries if:
NOTE:
It is necessary to set the date and time for the unit every time batteries are initially installed or replaced.
: Appears for User 1 : Appears for User 2 | |
Memory Average: Displays average of last 3 readings. | |
Pulse Symbol: Shows the heart rate per minute. | |
WHO Risk Category Indicator: See Blood Pressure Standard section. | |
Low Battery Symbol: Appears when batteries should be replaced. | |
Irregular Heartbeat Detector: See below for more information. | |
: Appears when cuff inflates : Appears when cuff deflates |
The appearance of the icon indicates that a pulse irregularity consistent with an irregular heartbeat was detected during measurement. Usually, this is not a cause for concern. However, if the symbol appears often, we recommend you seek medical advice. Please note that the device does not replace a cardiac examination, but serves to detect pulse irregularities at an early stage.
Movement, shaking, or talking during the measurement can result in pulse irregularities that may cause the appearance of this icon. Therefore, it is of great importance to not move or talk during measurement.
To determine the presence of an irregular heartbeat, the average of the heartbeat intervals is calculated with the first 3 normal effective heartbeat values. It is important to note that the average is not a strict mathematical averaging of all recorded intervals. At least 3 beats with 25% or greater difference from the average heartbeat interval will generate the icon on the screen.
Please note:
To use the right arm, you must position the artery symbol " " over the main artery. Locate the main artery by pressing with 2 fingers approximately 1" (2cm) above the bend of your elbow on the inside of your right arm. Identify where the pulse can be felt the strongest. This is your main artery.
DURING MEASUREMENT
NOTE:
This monitor can be used by 2 individuals. Each user can store up to 90 measurements.
From the power display off, press and hold down either the M1 or M2 button until the display shows CLr. This indicates that all measurements have been erased.
If any abnormality arises during use, please check the following points.
Problem | Probable Cause | Recommended Action |
Nothing appears in the display even when the power is turned on. | Batteries are drained. | Replace all batteries with new ones. |
Batteries are not correctly aligned with terminals. | Reinsert batteries in the correct position. | |
Low battery symbol appears. | Batteries are drained. | Replace all batteries with new ones. |
In colder temperatures, batteries have weaker electrical charges. | Warm up the batteries, or use the device in a warmer setting. | |
Device operation time is inconsistent. | Different battery brands have different life spans. | Use alkaline batteries and replace all batteries at the same time with the same brand of batteries. |
No reading after measurement. | Batteries are drained. | Replace all batteries with new ones. |
Suspicious blood pressure results. | Cuff positioned improperly. | Adjust patient and arm cuff to measure. |
Blood pressure naturally varies throughout the day. | Rest a while, relax, and measure again. | |
Suspicious heart rate results. | Body movement during device use. | Refrain from moving during measurement. |
Measurement shortly after exercise or exposure to the outdoors. | Do not take measurements after exercise or coming back from the outdoors. | |
Power switches off automatically. | System design. | Push the POWER button again, and then begin measure. |
During measuring, air reinflates. | It could be a normal action. If the user's blood pressure is higher than the initial pressure value, the device automatically pumps to a higher pressure by 40mmHg each time. | Relax, and try to take a measurement again. |
The arm cuff is not fastened properly. | Check that the arm cuff is fastened properly and retake the measurement. |
Note: If the unit still does not work, contact HoMedics Consumer Relations. Under no circumstance should you disassemble or attempt to repair the unit by yourself. Contact information for HoMedics Consumer Relations can be found on the Warranty page.
Note: These specifications are subject to change without notice.
Err Code | Meaning | Corrective Action |
Err 0 | Could not detect pulse. Too much body movement. | Reduce movement and retry measurement. |
Err 1 | Leakage in cuff pressure/inflation too low. | The arm cuff is not fastened properly. Reapply the cuff, and take a measurement again. |
Err 2 | Pressure fault, could not detect pulse. Too much body movement. | Rest a while, relax, and retry measurement. |
Err 3 | Could not detect pulse during deflation. Too much body movement. | The arm cuff is not fastened properly. Reapply the cuff, and take a measurement again. |
Err | Memory error. | Take off batteries to reboot the device, then take another measurement. |
Low batteries. | Replace all batteries with new ones. |
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download HoMedics BPA-O300 - Arm Blood Pressure Monitor Manual
Advertisement
Thank you! Your question has been received!
Need Assistance?
Do you have a question about the BPA-O300 that isn't answered in the manual? Leave your question here.