Troubleshooting; Specifications - HoMedics BPS-060-DDM Manual

Manual inflate blood pressure monitor
Table of Contents

Advertisement

Available languages

Available languages

TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on
Batteries have
when the
run down.
START/STOP
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement
Cuff has been placed
error symbol shown
incorrectly.
on display or the
blood pressure value
Did you talk or move
is displayed
during measurement?
excessively low
Shaking of the arm
(or high).
with the cuff on.
E1 error symbol
Air circuit abnormality.
shown on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol
Inflation pressure
shown on display
exceeding 300 mmHg.
E3 error symbol
Error determining
shown on display
measurement data.
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
29
CORRECTION
Replace them with
four new AA alkaline
batteries.
Re-insert the batteries
in the correct
positions.
Wrap the cuff
properly so that it is
positioned correctly.
Measure again.
Keep arm steady
during measurement.
Check cuff connection.
Measure again.
Turn the unit off, then
measure again.
Measure again.

SPECIFICATIONS

P P o o w w e e r r S S o o u u r r c c e e : :
Four 1.5V DC (AA) batteries
M M e e a a s s u u r r e e m m e e n n t t
Oscillometric
M M e e t t h h o o d d : :
M M e e a a s s u u r r e e m m e e n n t t
Pressure: 40~280 millimeters mercury (mmHg)
R R a a n n g g e e : :
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
A A c c c c u u r r a a c c y y : :
Pulse: ±5% of reading
P P r r e e s s s s u u r r e e S S e e n n s s o o r r : : Semi-conductor
I I n n f f l l a a t t i i o o n n : :
Manual Inflation
D D e e f f l l a a t t i i o o n n : :
Automatic Pressure release valve
M M e e m m o o r r y y C C a a p p a a c c i i t t y y : : 60 memories
A A u u t t o o - - s s h h u u t t - - o o f f f f : :
1 minute after last button operation
O O p p e e r r a a t t i i o o n n
Temperature: 50°F~104°F (10°C ~40°C)
E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: Less than 85% R.H.
S S t t o o r r a a g g e e
Temperature: -4°F~140°F (-20°C~60°C)
E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: Less than 85% R.H.
W W e e i i g g h h t t : :
0.43 lbs/193 grams (without batteries)
A A r r m m
Standard size adult arm cuff:
C C i i r r c c u u m m f f e e r r e e n n c c e e : :
9"–13" (23 – 33 cm)
5"(L) x 4.1"(W) x 2.7"(H)
D D i i m m e e n n s s i i o o n n s s : :
126mm(L) x 105mm(W) x 68mm(H)
Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the ingress
of water
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or nitrous
oxide.
Continuous operation with short-time loading.
Note: These specifications are subject to change without notice.