Troubleshooting; Specifications - HoMedics CareOne AHBPA-060 Manual

Table of Contents

Advertisement

Available languages

Available languages

TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Batteries have
Unit does not turn on
run down.
when the
START/STOP
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on dis-
incorrectly.
play or the blood pres-
sure value is displayed
excessively low
Did you talk or move
(or high).
during measurement?
Shaking of the arm with
the cuff on.
E1 error symbol shown
Air circuit abnormality.
on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on display
exceeding 300 mmHg.
E3 error symbol shown
Error determining
on display
measurement data.
CA Appears on
Calibration alert.
display
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair
the unit by yourself. Contact information for HoMedics Consumer
Relations Department can be found on the warranty page.
27
CORRECTION
Replace them with
four new AAA
alkaline batteries.
Re-insert the batteries
in the correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again. Keep
arm steady during
measurement.
Check cuff connection.
Measure again.
Switch the unit off, then
measure agaain.
Measure again.
Contact HoMedics
Consumer Relations
at 1-800-466-3342.

SPECIFICATIONS

Four 1.5V DC (AAA) batteries or
Power Source:
optional 6V DC AC adapter (not included)
Measurement
Oscillometric
Method:
Pressure: 40~280 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
50 memories for each User (100 total)
Auto-shut-off:
1 minute after last button operation
Operation
Temperature: 50°F~104°F (10°C ~40°C)
Environment:
Humidity: Less Than 85% RH
Temperature: -4°F~140°F (-20°C~60°C)
Storage Environment:
Humidity: Less Than 85% RH
Weight:
0.65 lb (295 g) (without batteries)
Arm Circumference:
Standard size cuff: 9"–13" (23-33 cm)
Large size cuff:13"–17" (33-43 cm)
5.6"(L) x 3.7"(W) x 2.8"(H)
Dimensions:
142mm(L) x 95mm(W) x 71mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
Note: These specifications are subject to change without notice.
28

Advertisement

Chapters

Table of Contents
loading

Table of Contents