Chapter 3 Replacing Assemblies
Removing the Rear Bezel
Figure 3-5. Top Rear Shroud Screws
4
Turn the Counter upside-down, and remove the three 4-40 x 3/16-inch
countersunk shroud screws on the bottom of the Counter near the back
edge of the shroud (see Figure 3-6).
Figure 3-6. Bottom Rear Shroud Screws
5
Carefully work the rear bezel out of the back end of the shroud by
alternately pulling opposite corners of the bezel away from the Counter a
small amount at a time. Do not use tools to pry the bezel out of the shroud,
since this could damage it.