Chapter 1 Performance Tests
Test 2: Channel 1 Frequency Sensitivity
Agilent 3325A/B Synthesizer
Reference
Output
(Rear Panel)
Agilent
53150A/51A/52A
Counter
Figure 1-3. 10 Hz to 20 MHz Frequency Sensitivity Test Setup
2
Verify that the Counter is still set to use an external reference signal
(refer to steps 5 through 10 on page 1-13).
3
Set the output of the Synthesizer to the first frequency specified for
Test 2a in the Performance Test Record on page 1-30 (10 Hz).
4
Set the power output of the Synthesizer to the first power value specified
for Test 2a in the Performance Test Record on page 1-30 (–14.9 dBm)
(40 mV).
5
Record the frequency value read on the Counter (±1 count) for Test 2a in
the Performance Test Record on page 1-30.
6
Repeat steps 3, 4, and 5 for each of the remaining frequencies specified for
Test 2a. Don't forget to change the power level when performing the
1 KHz test (–19.2 dBm) (25 mV).
7
If all of the frequency values you recorded for Test 2a are correct, mark
Pass in the Performance Test Record for Test 2a. If any of the frequency
readings you recorded are incorrect, mark Fail in the Performance Test
Record.
Assembly Level Service Guide
To Reference 10 MHz
Output
Connector (Rear Panel)
Channel 1
1-17