Chapter 1 Performance Tests
Test 1: External Timebase Input Test
To Reference 10 MHz
Connector
(Rear Panel)
Agilent
53150A/51A/52A
Counter
Figure 1-2. External Timebase Test Setup
4
Cycle the POWER button to preset the Counter.
5
Press the Shift key on the Counter's front panel.
6
Press the Menu key on the Counter's front panel.
7
If necessary, use the up- and down-arrow keys to cycle the display until it
reads REF OSC> INT.
8
Press the right-arrow key once. INT begins to flash off and on.
9
Press the up- or down-arrow key once. INT changes to EXT.
10
Press the Enter key.
11
Press the Chan Select key once to select the CHANNEL 1 input connector.
Assembly Level Service Guide
Agilent 3325A/B Synthesizer
Channel 1
Agilent 3325A/B Synthesizer
Output
Output
1-13