Long Term Tests - Fluke SigmaPace 1000 Operator's Manual

External pacemaker analyzer
Hide thumbs Also See for SigmaPace 1000:
Table of Contents

Advertisement

SigmaPace 1000
Operators Manual

Long Term Tests

Tests are preformed for only one channel at a time. The overall stability of the
pacemaker's pulse rate and amplitude are monitored continuously over an
extended period of time. The user enters the desired target values for these
two parameters into the Analyzer. This program flexibility allows the user to
enter either the actual test pacemaker values or the selected rate and amplitude
settings.
Deviation values are independently selectable from 2 % to 20 % in 1 %
increments. The Analyzer compares these incoming pacer pulses to the target
values. When the long-term test is complete, output pulses from the
pacemaker that were not within the selected limits may be reviewed.
F1
F2
F3
F4
eyr035.eps
1. From the SELECT TEST menu, press F3 LONG TERM.