Dc Load Test - Fluke SigmaPace 1000 Operator's Manual

External pacemaker analyzer
Hide thumbs Also See for SigmaPace 1000:
Table of Contents

Advertisement

SigmaPace 1000
Operators Manual

DC Load Test

1. From the SELECT TEST menu, press F2 DC LOAD.
2. Make necessary connections, as follows:
a.
Plug the supplied 9 V dc Load Test Cable into the matching
connector on the left side of the Analyzer.
b. A 9 V dc Load Test Cable (Fluke Biomedical Part# 2201074) is
supplied as a standard accessory with the Analyzer. It is designed
specifically for the testing of Medtronic models 5388 Dual Channel
and 5348 Single Channel transvenous pacemakers.
4-34
F1
F2
F1
F2
F3
F4
F3
F4
eyr034.eps
eyr083.eps

Advertisement

Table of Contents
loading

Table of Contents