SigmaPace 1000
Operators Manual
DC Load Test
1. From the SELECT TEST menu, press F2 DC LOAD.
2. Make necessary connections, as follows:
a.
Plug the supplied 9 V dc Load Test Cable into the matching
connector on the left side of the Analyzer.
b. A 9 V dc Load Test Cable (Fluke Biomedical Part# 2201074) is
supplied as a standard accessory with the Analyzer. It is designed
specifically for the testing of Medtronic models 5388 Dual Channel
and 5348 Single Channel transvenous pacemakers.
4-34