SigmaPace 1000
Operators Manual
Transvenous
Pacemaker
Atrial Channel
ECG
Sensing
& Pacer
Control
Ventricular
Channel
Power
ON/OFF
Potentially harmful dc levels can be detected using the various test
configurations described below:
Static Tests (Pacemaker Power OFF):
These tests continuously measure dc leakage as follows:
•
Atrial Channel: dc leakage across the atrial channel 500 Ω test load.
•
Ventricular Channel: dc leakage across the ventricular channel 500 Ω
test load.
•
AV Channel: dc leakage across an auxiliary 500 Ω test load
connected to the +atrial and +ventricular outputs.
4-8
SIGMAPACE 1000
External Pacemaker Analyzer
L
A+
O
A
D
A-
S
L
O
A
D
S
Figure 4-2. DC Leakage Testing
L
A+
O
A
V+
D
V+
V-
Static &
Dynamic
M
DC
Leakage
eyr006.eps