Fluke SigmaPace 1000 Operator's Manual page 83

External pacemaker analyzer
Hide thumbs Also See for SigmaPace 1000:
Table of Contents

Advertisement

3. For each, pulse rate and amplitude, press F3 LIMIT UP or DOWN arrow
key to select the limit value.
4. Press F4 LOAD UP or DOWN arrow key to select the load.
5. Press F1 NEXT to advance the screen.
6. Press F3 AMP UP or DOWN arrow key to adjust the amplitude.
7. Press F4 RATE UP or DOWN arrow key to adjust the rate.
Target values for amplitude and rate must be entered for the test to
perform correctly.
8. Press F1 START to advance the screen and start the test.
Transcutaneous Pacemaker Testing
F1
F2
Note
Setup and Testing
F3
F4
3
eyr027.eps
3-29