4. Press F2 SELECT DRIVE< to toggle between the Overdrive and
Underdrive BPM rates.
5. Once a rate is selected, press F4 DRIVE RATE BPM UP or DOWN
arrow key to adjust that rate.
6. Press ESC to return to previous menu(s).
Demand Mode Pacemaker's Ability to Sense ECG
Activity
The two following, related tests, PRP and SRP, determine the demand mode
pacemaker's ability to sense ECG activity immediately following either a
paced event (PRP) or a sensed ECG event (SRP).
A transcutaneous pacemaker must be set to Demand operation for
these tests.
Transcutaneous Pacemaker Testing
F1
F2
Note
Setup and Testing
F3
F4
3
eyr018.eps
3-17