Fluke SigmaPace 1000 Operator's Manual page 75

External pacemaker analyzer
Hide thumbs Also See for SigmaPace 1000:
Table of Contents

Advertisement

Transcutaneous Pacemaker Testing
3
Setup and Testing
2. Press the F4 LOAD UP or DOWN arrow key to select the load.
3. Press F1 NEXT to advance the screen for waveform selection.
F1
F2
F3
F4
eyr062.eps
4. Press F3 UP or DOWN arrow key to select waveform type.
5. Press F4 UP or DOWN arrow key to select waveform width.
6. Press F1 NEXT for amplitude and polarity selection.
3-21